Why the study?
Does device programming based on the 'fusion band' ECG morphology improve mechanical synchrony in patients with heart failure, sinus rhythm, and LBBB treated with single-site left ventricular pacing?
Does device programming based on the 'fusion band' ECG morphology improve mechanical synchrony in patients with heart failure, sinus rhythm, and LBBB treated with single-site left ventricular pacing?
Using ECG criteria based on the fusion band may offer a simple and rapid method to optimize resynchronization therapy in patients with heart failure, sinus rhythm, and LBBB.
Answers are grounded in this paper and cite their sources.
May support fusion band optimization in this pilot; leaves open whether it improves outcomes in randomized trials.
BACKGROUND: Electrical fusion between left ventricular pacing and spontaneous right ventricular activation is considered the key to resynchronisation in sinus rhythm patients treated with single-site left ventricular pacing. AIM: Use of QRS morphology to optimize device programming in patients with heart failure (HF), sinus rhythm (SR), left bundle branch block (LBBB), treated with single-site left ventricular pacing. METHODS AND RESULTS: We defined the "fusion band" (FB) as the range of AV intervals within which surface ECG showed an intermediate morphology between the native LBBB and the fully paced right bundle branch block patterns.Twenty-four patients were enrolled. Echo-derived parameters were collected in the FB and compared with the basal LBBB condition. Velocity time integral and ejection time did not improve significantly. Diastolic filling time, ejection fraction and myocardial performance index showed a statistically significant improvement in the FB. Interventricular delay and mitral regurgitation progressively and significantly decreased as AV delay shortened in the FB. The tissue Doppler asynchrony index (Ts-SD-12-ejection) showed a non significant decreasing trend in the FB. The indications provided by the tested parameters were mostly concordant in that part of the FB corresponding to the shortest AV intervals. CONCLUSION: Using ECG criteria based on the FB may constitute an attractive option for a safe, simple and rapid optimization of resynchronization therapy in patients with HF, SR and LBBB.
No takes yet. Share an insight, caveat, or question.
Gianfranchi et al. (2008) studied this question.