Answers are grounded in this paper and cite their sources.
A novel dynamically reconfigurable ECG analog front-end reduces power consumption by ~2.5× without compromising anomaly detection or feature extraction.
This paper presents a reconfigurable electrocardiogram (ECG) analog front-end (AFE) exploiting bio-signals' inherent low activity and quasi-periodicity to reduce power consumption. This is realized by an agile, on-the-fly dynamic noise-power trade-off performed over specific cardiac cycle regions guided by a least mean squares (LMS)-based adaptive predictor leading to ∼2.5× data-dependent power savings. Implemented in 65 nm CMOS, the AFE has tunable performance exhibiting an input-referred noise ranging from 2.38 – 3.64μ V_rmswhile consuming 307 – 769 nW from a 0.8 V supply. A comprehensive system performance verification was performed using ECG records from standard databases to establish the feasibility of the proposed predictor-based approach for power savings without compromising the system's anomaly detection capability or ability to extract pristine ECG features.
No takes yet. Share an insight, caveat, or question.
Mondal et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: