Key result
An ultra-low-power analog front end achieved a noise-efficient-factor of 2.4 with 4 μW power consumption and an input-reference noise of 0.39 μVrms, enabling clear P-wave acquisition for ECG.
Why the study?
The study aimed to develop an ultra-low-power low-noise analog front end for long-term ECG monitoring capable of capturing clear P-waves for clinical diagnosis.
A novel ultra-low-power analog front end demonstrates high-quality ECG signal acquisition with exceptionally low noise and power consumption, suitable for long-term monitoring.
Answers are grounded in this paper and cite their sources.
May enable extended wearable ECG monitoring; leaves open clinical validation in patients.
An ultra-low-power low-noise analog front end (AFE) is presented in this work, aiming for long-term ECG with clear P-waves for clinical diagnose. The chopper amplifier with passive noise filter and PWM based offset cancellation results in an input-reference noise of 0.39 μVrms, which shows 3.7X noise improvements among the state-of-the-art designs. With the digital offset cancellation by the pulse-width modulation wave, the AFE achieves a low input-referred dc offset of 0.4 μV among 9 tested chips and a low drift under 30 μV. A dynamic scale ADC with low-power comparator strategy prevents the instability and signal-loss, achieving an SFDR of 71.6 dB. The proposed AFE achieves a noise-efficient-factor (NEF) of 2.4 with a power consumption of 4 μW. The fabricated chip is demonstrated in a miniature prototype for long-term ECG monitoring application, presenting a clear ECG waveform with visible P-wave. The simultaneously ECG recording with a medical grade 12-lead ECG Holter shows the effective acquisition of the prototype, proofing the better noise performance.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2021) studied Long-term ECG monitoring (n=9). Ultra-low-power low-noise analog front end (AFE) vs. Medical grade 12-lead ECG Holter was evaluated on Noise-efficient-factor (NEF) and power consumption. An ultra-low-power analog front end achieved a noise-efficient-factor of 2.4 with 4 μW power consumption and an input-reference noise of 0.39 μVrms, enabling clear P-wave acquisition for ECG.