Why the study?
What are the adverse events associated with the implantation of a transvenous left ventricular pacing system in patients with severe heart failure and left bundle branch block?
What are the adverse events associated with the implantation of a transvenous left ventricular pacing system in patients with severe heart failure and left bundle branch block?
Transvenous left ventricular pacing via the coronary sinus is feasible in severe heart failure, though early lead dislodgement and threshold increases are frequent, highlighting the need for specialized leads.
Answers are grounded in this paper and cite their sources.
Early instability requires close initial monitoring after coronary sinus lead placement; leaves open need for dedicated lead designs in severe heart failure.
AIMS: Assessment of complications following implantation of transvenous ventricular electrodes to pace the left ventricle. METHODS AND RESULTS: Twenty-eight patients with severe cardiac failure and left bundle branch block were prospectively followed for adverse effects of implantation of a left ventricular transvenous pacing system. Immediate follow-up was associated with loss of left ventricular pacing in nine patients (32%). This was due to lead dislodgement in four cases (corrected by re-operation in three of these cases), and due to increased threshold in five cases (corrected by programming a higher pacing amplitude in all five cases, but with intermittent diaphragmatic contraction in one case). After 1 month, one patient died, one patient with severe coronary heart disease suffered a myocardial infarction, and left ventricular pacing was lost in two patients. Pericardial effusion, new significant ventricular arrhythmias or other adverse effects were not observed. After a mean follow-up of 16 +/- 9.2 months, pacing leads remained stable and no late complications related to the transvenous left ventricular epicardial pacing were observed. CONCLUSION: Placement of a permanent lead in a tributary of the coronary sinus is feasible without serious adverse effects during the first month. The only frequent adverse event was lead dislodgement; a finding which emphasizes the need for development of specially designed leads for this application.
No takes yet. Share an insight, caveat, or question.
V. Valls-Bertault (2001) studied this question.