A sensitive X-ray method is developed for precisely measuring stresses in thin films on single-crystal substrates, such as silicon wafers, and for determining strains in the surface layers of single crystals. A beam splitter cut from a large, dislocation-free single crystal produces two transmitted parallel beams of low angular divergency that penetrate the film-covered test crystal at two points. From the angular separation of the rocking curves at both points, the radius of curvature of the test crystal and, hence, the film stress are determined.
No takes yet. Share an insight, caveat, or question.
A. Bohg (1978) studied this question.