Key result
The Polar M600 optical heart rate sensor demonstrated 91.8% accuracy during cycle intervals but only 34.5% accuracy during circuit weight training compared to ECG.
Why the study?
Does the Polar M600 optical heart rate sensor accurately measure heart rate compared to ECG during various physical activities?
Observational (n=36)
Does the Polar M600 optical heart rate sensor accurately measure heart rate compared to ECG during various physical activities?
The Polar M600 optical heart rate sensor provides accurate heart rate measurements during steady-state exercise but is less accurate during rapid intensity changes and circuit weight training.
Answers are grounded in this paper and cite their sources.
Clinicians should interpret optical HR data cautiously during resistance exercise; leaves open the need for activity-specific validation of wearables.
PURPOSE: The purpose of this study was to evaluate the accuracy of the Polar M600 optical heart rate (OHR) sensor compared with ECG heart rate (HR) measurement during various physical activities. METHODS: Thirty-six subjects participated in a continuous 76-min testing session, which included rest, cycling warm-up, cycling intervals, circuit weight training, treadmill intervals, and recovery. HR was measured using a three-lead ECG configuration and a Polar M600 Sport Watch on the left wrist. Statistical analyses included OHR percent accuracy, mean difference, mean absolute error, Bland-Altman plots, and a repeated-measures generalized estimating equation design. OHR percent accuracy was calculated as the percentage of occurrences where OHR measurement was within and including ±5 bpm from the ECG HR value. RESULTS: Of the four exercise phases performed, the highest OHR percent accuracy was found during cycle intervals (91.8%), and the lowest OHR percent accuracy occurred during circuit weight training (34.5%). OHR percent accuracy improved steadily within exercise transitions during cycle intervals to a maximum of 98.5% and during treadmill intervals to a maximum of 89.0%. Lags in HR calculated by the Polar M600 OHR sensor existed in comparison to ECG HR, when exercise intensity changed until steady state occurred. There was a tendency for OHR underestimation during intensity increases and overestimation during intensity decreases. No statistically significant interaction effect with device was found in this sample on the basis of sex, body mass index, V˙O2max, skin type, or wrist size. CONCLUSIONS: The Polar M600 was accurate during periods of steady-state cycling, walking, jogging, and running, but less accurate during some exercise intensity changes, which may be attributed to factors related to total peripheral resistance changes and pulse pressure.
No takes yet. Share an insight, caveat, or question.
Horton et al. (2017) reported an observational. Polar M600 optical heart rate sensor vs. Three-lead ECG was evaluated on Optical heart rate percent accuracy (within ±5 bpm from ECG heart rate). The Polar M600 optical heart rate sensor demonstrated 91.8% accuracy during cycle intervals but only 34.5% accuracy during circuit weight training compared to ECG.