Key result
The Polar OH1 photoplethysmographic heart rate monitor showed high reliability for endurance sports compared to a chest belt, but lower reliability for sports with arm movements.
Why the study?
To assess the validity of the Polar OH1 photoplethysmographic heart rate monitor during various sports performed under ecological conditions.
Does the Polar OH1 armband accurately monitor heart rate and training load compared to the H7 chest belt in trained athletes during various sports?
Observational (n=70)
Does the Polar OH1 armband accurately monitor heart rate and training load compared to the H7 chest belt in trained athletes during various sports?
The Polar OH1 armband is a valid tool for monitoring instantaneous heart rate and training load, particularly in endurance sports, though accuracy is slightly reduced during sports with significant arm movement.
Answers are grounded in this paper and cite their sources.
Polar OH1 may suffice for endurance HR monitoring; leaves open validation needs in arm-movement sports.
This study assessed the validity of a photoplethysmographic heart rate (HR) monitor, the Polar OH1 in various sports performed in ecological conditions: running, cycling, soccer, kayaking, walking, tennis and fitness. Seventy trained athletes (56 males, 14 females) wore the Polar OH1 armband and the H7 chest belt during training. A total of 390 h and 38 min of recording were compared using a 20-bpm window to assess data quality, and Bland-Altman agreements and ICC analyses were used to test accuracy. Linear regression analyses evaluated the HR accuracy and correlation with skin tone. Training loads (TRIMPs) were compared for each session. Reliability was high for endurance sports (>99%) and lower for sports involving arm movements (92~95%). Biases were slightly negative for all sports, whereas widths of limits of agreement varied from 7-20bpm. Bland-Altman agreements were all under 5% except tennis, kayak and fitness. HR accuracy was positively correlated to skin tone (p<0.05). Finally, TRIMPs from the OH1 device were inferior to criterion's (except walking and soccer), within a 3% range from reference. Hence, OH1 represents a valid tool to monitor instantaneous HR and training load, especially for endurance sports.
No takes yet. Share an insight, caveat, or question.
Hermand et al. (2019) reported an observational. Polar OH1 armband vs. H7 chest belt was evaluated on Heart rate accuracy and reliability. The Polar OH1 photoplethysmographic heart rate monitor showed high reliability for endurance sports compared to a chest belt, but lower reliability for sports with arm movements.
Synapse has enriched 10 related papers. Consider them for comparative context: