Why the study?
Does the use of P wave amplitude histogram data improve the assessment of atrial sensitivity compared to a single measurement in patients with pacemakers?
Population
204 patients: 104 patients with a fixed atrial lead and a DDDR pacemaker and 100 patients with a single pass…
Comparison
P wave amplitude histogram feature vs Single measurement of the P wave amplitude at…
Design
Cohort
Authors
Loading...
Single P-wave measurement with 100% safety margin often fails reliable atrial sensing; leaves open whether histogram data improves programming in pacemaker patients.
Does the use of P wave amplitude histogram data improve the assessment of atrial sensitivity compared to a single measurement in patients with pacemakers?
P wave amplitude histograms provide a more reliable assessment of minimum P wave amplitudes than single measurements, potentially improving atrial sensing programming in pacemakers.
BOUTE et al. (1994) studied this question.