Key result
The UA-704 semi-automatic device achieved a British Hypertension Society grade A/A, with an average difference from mercury sphygmomanometer of -1.85±4.26 mmHg for systolic blood pressure.
Why the study?
Does the UA-704 semi-automatic device accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
Observational (n=85)
Does the UA-704 semi-automatic device accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
The UA-704 semi-automatic blood pressure device achieved an A/A grade according to the British Hypertension Society protocol and is recommended for home blood pressure monitoring.
Answers are grounded in this paper and cite their sources.
May support UA-704 for home monitoring; leaves open need for prospective validation in broader populations.
Objective To perform a clinical validation for a semi-automatic arm-type device UA-704 (A&D Company, Ltd., Tokyo, Japan) according to the British Hypertension Society protocol. Methods Eighty-five study participants (37 men) were included. The mean age was 50.1±17.0 years, systolic blood pressure range was 79–212 mmHg and diastolic blood pressure range was 43–118 mmHg. For each participant, three readings of the UA-704 were compared with simultaneous auscultatory measurements by two trained independent observers. The observers used a mercury calibrated sphygmomanometer and a dual stethoscope. The results were graded according to the 1993 British Hypertension Society protocol. Results The average difference between mercury sphygmomanometer and UA-704 readings was −1.85±4.26 mmHg (mean±SD) for systolic blood pressure and −1.44±3.97 mmHg for diastolic blood pressure. The proportions of values agreeing to within 5, 10 and 15 mmHg were 79.2%, 96.5% and 99.6% for systolic blood pressure and 86.7%, 96.9% and 99.6% for diastolic blood pressure for the observers and device (A/A grade for British Hypertension Society). Conclusions For an adult population, the UA-704 device for self-measurement of blood pressure achieved a British Hypertension Society grade A/A and therefore can be recommended for home blood pressure monitoring.
No takes yet. Share an insight, caveat, or question.
Kobalava et al. (2005) conducted an observational in Blood pressure measurement (n=85). UA-704 semi-automatic arm-type device vs. Mercury calibrated sphygmomanometer was evaluated on Difference between mercury sphygmomanometer and UA-704 readings (British Hypertension Society protocol) (Average difference -1.85±4.26 mmHg for SBP and -1.44±3.97 mmHg for DBP). The UA-704 semi-automatic device achieved a British Hypertension Society grade A/A, with an average difference from mercury sphygmomanometer of -1.85±4.26 mmHg for systolic blood pressure.