Does the A&D UA-705 semi-automatic monitor accurately measure blood pressure compared to a mercury standard in out-patients?
Does the A&D UA-705 semi-automatic monitor accurately measure blood pressure compared to a mercury standard in out-patients?
The A&D UA-705 semi-automatic blood pressure monitor is highly accurate and meets the British Hypertension Society standards for clinical use.
Answers are grounded in this paper and cite their sources.
May support UA-705 use in outpatients; leaves open broader validation across populations and settings.
OBJECTIVES: We tested the accuracy of the UA-705 blood pressure semi-automatic monitor. METHODS: Device evaluation was performed according to the modified British Hypertension Society protocol released in 1993. Eighty-five patients with characteristics outlined in the British Hypertension Society protocol were recruited among those attending our out-patient clinic. Sequential readings were taken for the main validation test. Outcome was classified according to the British Hypertension Society criteria, which are based on four zones of accuracy differing from the mercury standard by 5, 10 and 15 mmHg, or more. RESULTS: The mean blood pressure difference (+ or -1 SD) between device and observers was 0.4 mmHg (SD 7.7) for systolic blood pressure and 0.3 mmHg (SD 8.2) for diastolic blood pressure. Overall, 96% (observer 1) and 95% (observer 2) of readings between device and observers differed by 15 mmHg or less for both systolic and diastolic blood pressure. The device achieved a grade A for both systolic and diastolic blood pressure. CONCLUSIONS: This study shows that the A&D UA-705 device satisfies the British Hypertension Society standard for accuracy by achieving a grade A for both systolic and diastolic blood pressure.
No takes yet. Share an insight, caveat, or question.
Verdecchia et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: