Key result
Expansion dislodgement of the cryoballoon during atrial fibrillation ablation occurred in 22.7% of applications by intracardiac echocardiography and 32% by dielectric imaging.
Why the study?
Intraballoon pressure surges during cryoballoon ablation may cause balloon dislodgement and loss of PV occlusion, which can lead to incomplete isolation but often goes undetected without continuous monitoring.
What is the incidence of cryoballoon expansion dislodgement and loss of pulmonary vein occlusion during the initial phase of freezing in patients undergoing ablation for atrial fibrillation?
Observational (n=15)
What is the incidence of cryoballoon expansion dislodgement and loss of pulmonary vein occlusion during the initial phase of freezing in patients undergoing ablation for atrial fibrillation?
Cryoballoon expansion dislodgement occurs in approximately one-fourth to one-third of applications during pulmonary vein isolation, representing a frequent and potentially underappreciated cause of incomplete isolation.
Answers are grounded in this paper and cite their sources.
Continuous pressure monitoring may detect dislodgement and reduce incomplete PVI; leaves open whether it improves long-term outcomes.
BACKGROUND: Cryoballoon ablation for atrial fibrillation (AF) requires adequate contact between the pulmonary vein (PV) antrum and cryoballoon. The surge of intraballoon pressure during the initial phase of ablation may change the balloon's shape and compliance, resulting in balloon dislodgement and loss of PV occlusion. Without continuous monitoring, this phenomenon is often undetected but can be associated with incomplete PV isolation (PVI). METHODS: Primary cryoablation of AF was performed in 15 patients. PV occlusion status pre- and post-freezing were analyzed with intracardiac echocardiography (ICE) and dielectric imaging-based occlusion tool (DIOT) to calculate the incidence of expansion dislodgement of cryoballoon. RESULTS: A total of 105 cryoablation applications were performed on 57 veins, including three common ostiums of left pulmonary veins. In the evaluation of PV occlusion, both modalities reported consistent results in 86.7% of the assessments. Despite complete PV occlusion before ablation, peri-balloon leak after initiation of freezing was detected by ICE in 5/22 (22.7%) applications and by DIOT in 8/25 (32%) applications. CONCLUSION: Incidence of expansion dislodgement of the cryoballoon was detected in one-fourth to one-third of cryoablation applications depending on the imaging modality used, which was clinically frequent and significant.
No takes yet. Share an insight, caveat, or question.
Liao et al. (2023) conducted an observational in Atrial fibrillation (n=15). Cryoballoon ablation was evaluated on Peri-balloon leak after initiation of freezing (expansion dislodgement). Expansion dislodgement of the cryoballoon during atrial fibrillation ablation occurred in 22.7% of applications by intracardiac echocardiography and 32% by dielectric imaging.
Synapse has enriched 10 related papers. Consider them for comparative context: