Key result
The multielectrode radiofrequency-balloon catheter achieved successful pulmonary vein isolation in 100% of targeted veins, with 73.2% isolated by a single application and a median procedure time of 62.5 minutes.
Why the study?
The operator learning curve and procedural outcomes during the clinical implementation of the novel multielectrode RF-balloon catheter for PVI in AF had not been evaluated at a high-volume center.
Does the multielectrode radiofrequency balloon catheter provide efficient and safe pulmonary vein isolation in patients with atrial fibrillation during the operator learning curve?
Observational (n=40)
No
Does the multielectrode radiofrequency balloon catheter provide efficient and safe pulmonary vein isolation in patients with atrial fibrillation during the operator learning curve?
Rapid implementation of a 'single shot' multielectrode RF-balloon catheter for pulmonary vein isolation can be achieved with high acute procedural efficacy, short procedure times, and safety at a high-volume center.
Background: The novel multielectrode radiofrequency (RF) balloon catheter (HELIOSTAR™, Biosense Webster) is a new technology for pulmonary vein isolation (PVI) in atrial fibrillation (AF), combining RF-ablation and 3D-mapping visualization with the concept of a "single-shot"-ablation device. This study evaluates the operator learning curve und procedural outcome during implementation of the multielectrode RF-balloon at a high-volume center. Methods: The first 40 patients undergoing PVI by multielectrode RF-balloon catheter at Heidelberg University Hospital were included in this prospective study. Procedural outcome was analyzed over the course of increasing experience with the device. Results: 157/157 pulmonary veins (PVs) were successfully isolated with the RF-balloon catheter, in 73.2% by a single RF-application. Median time to isolation (TTI) was 11.0 s (Q1 = 8.0 s; Q3 = 13.8 s). Median procedure time was 62.5 min (Q1 = 50.0 min; Q3 = 70.5 min). LA-dwell time was 28.5 min (Q1 = 23.3 min; Q3 = 36.5 min). Median fluoroscopy duration was 11.6 min (Q1 = 10.1 min; Q3 = 13.7 min). No serious procedure-related complications were observed, apart from one case of unclear, post-procedural acute-on-chronic kidney injury. With increasing operator experience, an additional reduction in procedure duration was observed. Conclusion: Rapid implementation of a "single shot"-ablation device combining RF-ablation and 3D-mapping can be achieved with high acute procedural efficacy and safety at a high-volume center. Previous experience with "single-shot" ablation devices may be advantageous for time-efficient introduction of the novel RF-balloon catheter into clinical practice. Clinical Trial Registration: ClinicalTrials.gov; Identifier NCT0560361.
No takes yet. Share an insight, caveat, or question.
Zylla et al. (2023) conducted an observational in Atrial fibrillation (n=40). Multielectrode radiofrequency-balloon catheter (HELIOSTAR) was evaluated on Successful pulmonary vein isolation. The multielectrode radiofrequency-balloon catheter achieved successful pulmonary vein isolation in 100% of targeted veins, with 73.2% isolated by a single application and a median procedure time of 62.5 minutes.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: