Why the study?
Can a new continuous blood pressure measurement method based on a pulse image sensor accurately estimate systolic and diastolic blood pressure?
Can a new continuous blood pressure measurement method based on a pulse image sensor accurately estimate systolic and diastolic blood pressure?
A novel continuous blood pressure measurement method using a pulse image sensor demonstrates accuracy meeting AAMI requirements in a simulation platform.
Answers are grounded in this paper and cite their sources.
Supports technical feasibility of pulse image sensor BP monitoring; leaves open clinical validation before adoption.
In order to find a new measurement method for continuous blood pressure (BP), pulse image sensor, pressure sensor and data acquisition card MP425 were used to build the BP simulation platform of multi-information synchronous acquisition, including dynamic pulse images, pressure on the wrist (POW) and BP signals. Multi-point pulse wave (PW) amplitudes of film grid on the pulse image sensor were extracted from the dynamic pulse images by grid area method. Diastolic BP (DBP) and systolic BP (SBP) were estimated by multi-linear regression analysis. The results show that the maximum error of SBP (mmHg) and DBP (mmHg) satisfy the AAMI requirements (mmHg).