This is the authors' abstract. We don't add key points for this paper.
In this work, binary transition-metal oxide nanomaterial NiCo 2 O 4 @NiMoO 4 nanotip arrays (NAs) with core–shell heterostructures were directly coated on carbon cloth (CC) by facial hydrothermal treatments. The prepared NiCo 2 O 4 @NiMoO 4 NAs/CC was used as an advanced binder-free, flexible, and nonenzymatic electrode material for glucose (Glu) detection directly. Thanks to the synergistic effect of fast electron transport and penetration, the NiCo 2 O 4 @NiMoO 4 NAs/CC heterojunction sensor presented superior Glu sensing properties and exhibited a low limit of detection (LOD) of 0.24 μM with an ultrahigh sensitivity of 9557 μA mM –1 cm –2 and, particularly, a fast response time (2.1 s) besides good selectivity or stability. Moreover, good accuracy and reproducibility can also be obtained for Glu determination in the human serum sample. The above conclusions demonstrate that the construction of the NiCo 2 O 4 @NiMoO 4 NAs/CC heterojunction is conducive to the improvement of nonenzymatic Glu detection.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: