Why the study?
Quantifying distinct factors influencing local impedance measurements remains challenging due to lack of ground truth knowledge and superposition of effects in clinical settings.
The developed in silico environment accurately simulates clinical scenarios for local impedance measurements, providing a tool to assist interpretation of human measurements and support future ablation catheter development.
Answers are grounded in this paper and cite their sources.
Simulation quantifies local impedance factors in ablation; hypothesis-generating and requires in vivo validation before clinical use.
BACKGROUND: Electrical impedance measurements have become an accepted tool for monitoring intracardiac radio frequency ablation. Recently, the long-established generator impedance was joined by novel local impedance measurement capabilities with all electrical circuit terminals being accommodated within the catheter. OBJECTIVE: This work aims at in silico quantification of distinct influencing factors that have remained challenges due to the lack of ground truth knowledge and the superposition of effects in clinical settings. METHODS: We introduced a highly detailed in silico model of two local impedance enabled catheters, namely IntellaNav MiFi OI and IntellaNav Stablepoint, embedded in a series of clinically relevant environments. Assigning material and frequency specific conductivities and subsequently calculating the spread of the electrical field with the finite element method yielded in silico local impedances. The in silico model was validated by comparison to in vitro measurements of standardized sodium chloride solutions. We then investigated the effect of the withdrawal of the catheter into the transseptal sheath, catheter-tissue interaction, insertion of the catheter into pulmonary veins, and catheter irrigation. RESULTS: All simulated setups were in line with in vitro experiments and in human measurements and gave detailed insight into determinants of local impedance changes as well as the relation between values measured with two different devices. CONCLUSION: The in silico environment proved to be capable of resembling clinical scenarios and quantifying local impedance changes. SIGNIFICANCE: The tool can assists the interpretation of measurements in humans and has the potential to support future catheter development.
No takes yet. Share an insight, caveat, or question.
Unger et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: