Key result
Ablation with the Intellanav MiFi OI catheter showed a larger median local impedance drop for successful versus unsuccessful lesions in the left ventricle.
Why the study?
Does the Intellanav MiFi OI catheter provide local impedance data that correlates with successful lesion formation in patients undergoing ablation?
Observational (n=31)
Does the Intellanav MiFi OI catheter provide local impedance data that correlates with successful lesion formation in patients undergoing ablation?
The novel Intellanav MiFi OI catheter provides reproducible local impedance recordings that correlate with successful lesion formation during ablation.
Answers are grounded in this paper and cite their sources.
Initial experience supports feasibility of impedance-guided mapping; leaves open clinical outcome benefits pending larger trials.
INTRODUCTION: Successful catheter ablation is limited by both poor spatial resolution of abnormal local signals and inability to deliver an effective lesion due to poor tissue contact. We report first worldwide use of the Intellanav MiFi OI catheter (Boston Scientific), providing ultra-high density mapping and incorporating a "DirectSense" algorithm to measure local tissue impedance (LI). METHODS AND RESULTS: 31 patients (65±6 years, 20 male) underwent ablation. LI from the catheter, generator impedance (GI) and maximum electrogram amplitude were recorded in the blood pool, and in regions from healthy to dense scar before, during and after ablation. The catheter demonstrated clear nearfield signal where standard bipolar recordings included farfield signal. LI was lower in dense scar than either healthy tissue or blood pool, and demonstrated an exponential relationship with maximum electrogram amplitude. Maximum LI drop on ablation linearly correlated with initial LI. The median LI drop for successful lesions, resulting in lack of local tissue capture, was 16.0Ω (12.1-19.8 Ω) for LV and 14.6 Ω (10.0-18.3 Ω) for LA, which was larger than for unsuccessful lesions (LV: 9.4 Ω [5.4-15.6 Ω] P = 0.001; LA: 6.8 Ω [4.7-13.0 Ω], P = 0.049). LI percentage drop was also significantly larger for successful than unsuccessful lesions (LV: 17.1 Ω [14.0-19.6 Ω] vs. 10.6 Ω (7.1-16.5 Ω) P = 0.002; LA: 14.2 Ω [10.8-19.5 Ω] vs. 7.5Ω [5.1-11.0 Ω], P = 0.005). CONCLUSION: This novel catheter gives reproducible recordings of local impedance, which are dependent on scar level. Absolute LI drop, and also percentage drop, on ablation may give an indication of tissue contact and subsequent effective lesion formation.
No takes yet. Share an insight, caveat, or question.
Martin et al. (2018) conducted an observational in Arrhythmia requiring catheter ablation (n=31). Intellanav MiFi OI catheter vs. Unsuccessful lesions was evaluated on Median local impedance drop for successful vs unsuccessful lesions in the left ventricle (p=0.001). Ablation with the Intellanav MiFi OI catheter showed a larger median local impedance drop for successful versus unsuccessful lesions in the left ventricle.
Synapse has enriched 10 related papers. Consider them for comparative context: