Why the study?
Does radiofrequency catheter ablation using the CARTO UNIVU system reduce fluoroscopy exposure and catheter use compared to CARTO 3 in children with left-sided accessory pathways?
Does radiofrequency catheter ablation using the CARTO UNIVU system reduce fluoroscopy exposure and catheter use compared to CARTO 3 in children with left-sided accessory pathways?
The use of a fluoroscopy-integrated 3D-mapping system (CARTO UNIVU) for left-sided accessory pathway ablation in children is safe, highly effective, and significantly reduces radiation exposure compared to standard 3D mapping.
Answers are grounded in this paper and cite their sources.
May reduce fluoroscopy and catheters in pediatric left AP ablation; leaves open need for randomized confirmation.
Introduction: Advances in 3D electroanatomic-mapping technologies have resulted in a safe and effective profile of radiofrequency (RF) catheter ablation. The aim of this study was to evaluate a different catheter ablation approach in patients with left-sided accessory pathways (APs). Methods and results: From January 2015 to December 2015, 30 patients (median age 11 years, median weight 45 kg) with manifest or concealed left-sided APs underwent RF catheter ablation with a new protocol. All procedures were performed with the CARTO UNIVU™ system, integrating electroanatomic maps with fluoroscopic views. A 7 Fr ablation catheter was inserted into the right femoral vein and advanced into the right atrium. Geometrical reconstruction and activation map of the right atrium, tricuspid annulus, and coronary sinus were acquired. The ablation catheter was then inserted into the left femoral artery and advanced through the aorta and aortic valve, creating an activation map of the mitral annulus. Catheter ablation was targeted to the site of the earliest activation. No complications occurred. The median procedure and fluoroscopy times were 130 min and 6 s, with a median fluoroscopy dose 0.5 mGy. An average of two catheters was used. Long-term success rate was 97% (29/30) at a median follow-up of 9.6 months. This approach reduced fluoroscopy time, dose and number of catheters used compared with manifest or concealed left-sided AP ablation using CARTO 3™ (P < 0.05). Conclusions: This new ablation protocol seems to be promising in reducing fluoroscopy exposure and number of catheters used during left-sided AP ablation in children.
No takes yet. Share an insight, caveat, or question.
Drago et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: