Why the study?
Does a modular system for detecting plethysmographic blood volume pulse amplitude accurately correlate with manual trough to peak ruler measurements?
Does a modular system for detecting plethysmographic blood volume pulse amplitude accurately correlate with manual trough to peak ruler measurements?
The described modular system provides a valid, simpler, and faster alternative to manual hand-scoring for quantifying plethysmographic blood volume pulse amplitude.
Answers are grounded in this paper and cite their sources.
Modular system may simplify BVP assessment; leaves open prospective validation before routine clinical adoption.
A modular system for detecting and displaying the maximum amplitude of a plethysmographic blood volume pulse is described. By utilizing a peak detector and supporting logic, the height of the pulse wave for each cardiac cycle is displayed as an analogue step function similar in appearance to the output of a cardiotachometer. This output is simpler and faster to quantify than the typical hand‐scoring of blood volume pulse amplitude and is easily adaptable as a feedback medium for pulse wave information. The validity of the new method was ascertained by comparing its output with trough to peak ruler measurements of blood volume pulse during a variety of events. For each stimulus condition correlations between the two methods of scoring ranged from r = .85 to .99 with an average of r = .93 while within subjects correlations ranged from r = .87 to .99 with an average of r = .95.
No takes yet. Share an insight, caveat, or question.
Cohen et al. (1979) studied this question.