Key result
Deep neural network models for the ARTSENS A-mode ultrasound device measured elastic modulus with an error of 9.3% and a Pearson correlation coefficient of 0.94.
Why the study?
Carotid arterial stiffness is an early cardiovascular disease risk marker, but measuring elastic modulus with the image-free ARTSENS device requires automating carotid detection, wall localization, and inner lumen diameter measurement.
Deep neural network models can accurately automate carotid detection and arterial stiffness measurement using image-free A-mode ultrasound.
Supports portable elastic modulus assessment via DNN-enhanced A-mode ultrasound; leaves open prospective clinical validation before practice adoption.
Arterial stiffness (AS) of the carotid artery is an early marker of stratifying cardiovascular disease risk. This article aims to improve the performance of ARTSENS, a noninvasive A-mode ultrasound-based device for measuring AS. The primary objective of ARTSENS is to enable the measurement of elastic modulus using A-Mode ultrasound and blood pressure. As this device is image-free, there is a need to automate: 1) carotid detection; 2) wall localization; and 3) inner lumen diameter measurement. This has been performed using conventional signal processing methods in some of the earlier works in this domain. In this article, deep neural network (DNN) models are employed to perform the above three tasks. The DNNs were trained over data acquired from 82 subjects at two different medical centers. Ground-truth labeling was performed by a trained operator using corresponding measurements from the state-of-the-art Aloka e-Tracking system. All three DNN models had significantly lower errors compared to earlier signal processing methods and could perform their measurements using a single A-Mode frame. Using the DNNs, two different machine learning pipelines have been proposed here to measure the elastic modulus; the best among them could achieve an error of 9.3% with the Pearson correlation coefficient of 0.94 ( ). The models were tested on Raspberry Pi and Jetson Nano single board computers to demonstrate real-time processing on low computational resources.
No takes yet. Share an insight, caveat, or question.
Sahani et al. (2021) studied Arterial stiffness (n=82). Deep neural network (DNN) models vs. Conventional signal processing methods was evaluated on Measurement of elastic modulus (Pearson correlation 0.94). Deep neural network models for the ARTSENS A-mode ultrasound device measured elastic modulus with an error of 9.3% and a Pearson correlation coefficient of 0.94.