Key result
An anaeroid manometer and an automated device both met AAMI criteria (<5 mmHg difference) compared to a mercury standard, though the automated device overestimated SBP by 2.12 mmHg.
Why the study?
Do an anaeroid manometer and an automated oscillometric device provide accurate blood pressure measurements compared to a traditional mercury sphygmomanometer in human participants?
Cross-Sectional (n=512)
Do an anaeroid manometer and an automated oscillometric device provide accurate blood pressure measurements compared to a traditional mercury sphygmomanometer in human participants?
Both an anaeroid manometer and an automated oscillometric device met AAMI criteria for accuracy and may serve as viable replacements for traditional mercury sphygmomanometers.
Answers are grounded in this paper and cite their sources.
May support anaeroid and automated devices as mercury alternatives; leaves open prospective validation in diverse populations.
OBJECTIVE: Alternatives to the traditional, but possibly toxic mercury sphygmomanometer are needed for accurate blood pressure measurements in the medical workplace. We compared the performance of two commercially available potential replacements for the mercury column; an anaeroid manometer (Baum & Co) and an automated oscillometric device (Omron HEM-907), using the mercury sphygmomanometer as a standard, in the same participants. METHODS: Two independent observers performed simultaneous triplicate blood pressure readings for 512 participants. The average difference and standard deviation of the difference comparing the mercury column vs. the anaeroid and automated devices were calculated for each of the three paired systolic and diastolic blood pressure readings. RESULTS: Both devices met the Association for the Advancement of Medical Instrumentation criteria for sphygmomanometers (<5 mmHg average difference, <8 mmHg standard deviation of the difference) for all three readings. Compared with the mercury standard, there were no significant differences (by paired t-test) with the anaeroid device (-0.83/0.73 mmHg, P=0.25/0.09), but the automated device slightly overestimated systolic blood pressure (by 2.12 mmHg, P=0.002) and underestimated diastolic blood pressure (by 2.36 mmHg, P=0.0002). The first reading was significantly higher and had a larger standard deviation than the second or third readings across all manometers. CONCLUSIONS: The automated device performed as well as an anaeroid manometer operated by well trained, experienced observers. The two alternative devices to the mercury sphygmomanometer examined in this study may be potential replacement devices for blood pressure measurement.
No takes yet. Share an insight, caveat, or question.
Elliott et al. (2007) conducted a cross-sectional in Blood pressure measurement (n=512). Anaeroid manometer and automated oscillometric device vs. Mercury sphygmomanometer was evaluated on Average difference and standard deviation of the difference in systolic and diastolic blood pressure readings. An anaeroid manometer and an automated device both met AAMI criteria (<5 mmHg difference) compared to a mercury standard, though the automated device overestimated SBP by 2.12 mmHg.