Why the study?
Under the Minamata Convention on Mercury, traditional gold-standard mercury sphygmomanometers are being replaced by automated oscillometric devices, requiring scientific evidence on whether automated devices can replace them in routine practice.
Do automated oscillometric devices provide comparable blood pressure measurements to mercury sphygmomanometers?
Do automated oscillometric devices provide comparable blood pressure measurements to mercury sphygmomanometers?
Automated oscillometric devices measure slightly lower blood pressure than mercury sphygmomanometers (within 5 mmHg), with significant heterogeneity depending on the device manufacturer.
Answers are grounded in this paper and cite their sources.
Oscillometric devices may underestimate BP; leaves open calibrated replacement of mercury sphygmomanometers pending validation standards.
OBJECTIVE: A mercury sphygmomanometer has been considered a gold standard for measuring blood pressure. However, by the Minamata Convention on Mercury, the traditional mercury sphygmomanometer is being replaced by an automated oscillometric device. This study aimed to provide scientific evidence to determine whether an automated oscillometric device can replace a mercury sphygmomanometer and if it is applicable in routine practice. METHODS: MEDLINE, EMBASE, the Cochrane Library, and CINAHL were searched on 4 May 2018. Studies comparing blood pressure measurements between automated oscillometric devices and mercury sphygmomanometers were included. Study characteristics were abstracted using the evidence table, and random-effects meta-analyses were conducted. RESULTS: Data were compiled from 24 studies comprising 47 759 subjects. The results of meta-analysis showed that automated oscillometric devices measured lower than mercury sphygmomanometers for both systolic blood pressure (mean differences -1.75 mmHg, 95% confidence intervals: -3.05 to -0.45, I = 91.0%) and diastolic blood pressure (mean differences -1.20 mmHg, 95% confidence intervals: -2.16 to -0.24, I = 95.0%). In sub-group analyses by manufacturer, BpTRU measured lower than the mercury sphygmomanometer and OMRON showed no difference compared to the mercury sphygmomanometer for both systolic and diastolic blood pressure, but the results differed depending on the devices. CONCLUSION: As a result of this review, the difference in blood pressure between the mercury sphygmomanometer and the automated oscillometric device was within 5 mmHg, but the heterogeneity between the studies was very high. The automated oscillometric devices showed differences in blood pressure results according to the manufacturer and product type.
No takes yet. Share an insight, caveat, or question.
Park et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: