Why the study?
TAV are symmetrically designed but often deploy in heavily calcified, noncircular conduits resulting in deformed prosthesis geometries.
Do deformed valve annulus configurations affect valve performance metrics in a computational model of a bioprosthetic valve?
Do deformed valve annulus configurations affect valve performance metrics in a computational model of a bioprosthetic valve?
Computational modeling demonstrates that under-expansion of bioprosthetic valves primarily increases stenosis, whereas noncircular deformation primarily increases regurgitation and mechanical stress.
Answers are grounded in this paper and cite their sources.
May increase leaflet stress in deformed annuli; leaves open impact on TAV durability.
Transcatheter aortic valves (TAV) are symmetrically designed, but they are often not deployed inside cylindrical conduits with circular cross-sectional areas. Many TAV patients have heavily calcified aortic valves, which often result in deformed prosthesis geometries after deployment. We investigated the effects of deformed valve annulus configurations on a surgical bioprosthetic valve as a model for TAV. We studied valve leaflet motions, stresses and strains, and analog hydrodynamic measures (using geometric methods), via finite element (FE) modeling. Two categories of annular deformations were created to approximate clinical observations: (1) noncircular annulus with valve area conserved, and (2) under-expansion (reduced area) compared to circular annulus. We found that under-expansion had more impact on increasing stenosis (with geometric orifice area metrics) than noncircularity, and that noncircularity had more impact on increasing regurgitation (with regurgitation orifice area metrics) than under-expansion. We found durability predictors (stress/strain) to be the highest in the commissure regions of noncircular configurations such as EllipMajor (noncircular and under-expansion areas). Other clinically relevant performance aspects such as leaflet kinematics and coaptation were also investigated with the noncircular configurations. This study provides a framework for choosing the most challenging TAV deformations for acute and long-term valve performance in the design and testing phase of device development.
No takes yet. Share an insight, caveat, or question.
Jafar et al. (2019) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: