Why the study?
Does pulse oximetry accurately measure oxygen saturation compared to co-oximetry in patients with septic shock and low systemic vascular resistance?
Does pulse oximetry accurately measure oxygen saturation compared to co-oximetry in patients with septic shock and low systemic vascular resistance?
Pulse oximetry slightly but significantly underestimates oxygen saturation in patients with septic shock and low systemic vascular resistance, likely due to sensing pulsatile venous flow.
Answers are grounded in this paper and cite their sources.
Caution advised interpreting pulse oximetry in low-SVR septic shock; leaves open need for prospective co-oximetry validation.
In order to assess the accuracy of pulse oximeters in patients with septic shock, we compared 80 paired readings of oxygen saturations taken from pulse oximeters and oxygen saturations obtained from co-oximetry in patients receiving intensive therapy with indwelling pulmonary artery flotation catheters. Comparison between groups with low or normal systemic vascular resistance indices showed a small (1.4%) but significant (p < 0.001) underreading of the saturation from the pulse oximeter in the presence of a low systemic vascular resistance. With normal or high systemic vascular resistance pulse oximeter readings correlated well with co-oximetry. We hypothesise that the main cause of this underreading is because the pulse oximeter is sensing pulsatile venous flow due to the opening of arteriovenous channels in the skin in septic states.
No takes yet. Share an insight, caveat, or question.
Secker et al. (1997) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: