Key result
Successful single-shot pulmonary vein isolation with a novel radiofrequency balloon catheter showed higher mean impedance drop than non-single-shot applications.
Why the study?
A novel multielectrode radiofrequency balloon catheter has been released for pulmonary vein isolation, but the impact of ablation parameters on single-shot isolation required evaluation.
Are impedance drop and temperature rise associated with successful single-shot pulmonary vein isolation using a novel radiofrequency balloon catheter in patients with atrial fibrillation?
Observational (n=105)
Yes
Are impedance drop and temperature rise associated with successful single-shot pulmonary vein isolation using a novel radiofrequency balloon catheter in patients with atrial fibrillation?
The novel radiofrequency balloon catheter achieves high acute pulmonary vein isolation success, with single-shot isolation being predictable by greater impedance drops and temperature rises during ablation.
Answers are grounded in this paper and cite their sources.
Greater impedance drop and temperature rise may predict single-shot success with this balloon; hypothesis-generating for guided ablation, needs prospective validation.
BACKGROUND: A novel multielectrode radiofrequency balloon (RFB) catheter has been released for pulmonary vein isolation (PVI). METHODS AND RESULTS: ) were included. 241/412 (58.5%) PVs were successfully isolated with a single shot (SS), with a time-to-isolation of 11.6±8 s. Total number of radiofrequency applications was 892 (mean 2.2/PV), resulting in successful isolation of 408/412 (99%) PVs at the end of the procedure. Mean electrodes' impedance drop was significantly higher in the SS-PVI compared with non-SS applications (21.5±6.6 vs. 18.6±6.5 Ohm). Concordantly, higher temperature rise was observed in the SS vs. non-SS applications (10.9±4.9℃ vs. 9.6±4.7℃). CONCLUSIONS: In this multicenter real-world study, mean impedance drop and temperature rise were associated with successful SS-PVI applying the novel RFB catheter. These parameters may help to guide efficient usage of the new RF balloon.
No takes yet. Share an insight, caveat, or question.
My et al. (2023) conducted an observational in Paroxysmal or persistent atrial fibrillation (n=105). Single-shot radiofrequency balloon catheter application vs. Non-single-shot application was evaluated on Mean impedance drop during radiofrequency application (p=<0.001). Successful single-shot pulmonary vein isolation with a novel radiofrequency balloon catheter showed higher mean impedance drop than non-single-shot applications.
Synapse has enriched 10 related papers. Consider them for comparative context: