Do SpaceLabs ambulatory non-invasive blood pressure recorders maintain measurement consistency over several years of use?
Do SpaceLabs ambulatory non-invasive blood pressure recorders maintain measurement consistency over several years of use?
SpaceLabs ambulatory blood pressure recorders maintain reliable measurement consistency over 6 years of regular use.
Answers are grounded in this paper and cite their sources.
Supports long-term use of these recorders without frequent recalibration; leaves open generalizability to other devices and populations.
OBJECTIVE: To assess the measurement consistency of SpaceLabs ambulatory recorders (Spacelabs, Washington, USA) that are in regular use. METHODS: A total of 14 SpaceLabs 90207 and one 90217 ambulatory recorders were tested for measurement consistency using the Dynatech CuffLink (Dynatech, Nevada, USA), a commercially available non-invasive blood pressure (NIBP) simulator. The NIBP recorders were tested at a range of pressures with 20 repeated determinations at a simulated 120/80 mmHg and five repeated determinations at simulated pressures of 80/50, 100/80, 150/100, 200/165 and 250/195 mmHg. Tests were carried out in 1998, 2002 and late 2003 or early 2004. CONCLUSIONS: All 15 SpaceLabs recorders measured consistently over the 6 years with 89.5% of the differences in average pressures, recorded by any particular device at each recorded pressure, less than 2 mmHg between successive test episodes. The maximum difference was 4.5 mmHg and 60.1% of the differences were less than 1 mmHg. The measurements for all devices were within the tolerances specified by the supplier for the device when tested with the simulator. Maintenance records also show that most devices required breakdown maintenance less than once every 3 years. The results show that the SpaceLabs devices maintain measurement consistency in the demanding conditions of ambulatory pressure recording over several years.
No takes yet. Share an insight, caveat, or question.
Amoore et al. (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: