Why the study?
A longer-arm clip design for transcatheter edge-to-edge repair of mitral and tricuspid valves could be beneficial in complex pathologies, but its hemodynamic effects and usability are unknown.
Does long-arm clip implantation improve hemodynamics in an ex-vivo porcine model of mitral and tricuspid regurgitation?
Does long-arm clip implantation improve hemodynamics in an ex-vivo porcine model of mitral and tricuspid regurgitation?
In an ex-vivo porcine model, a long-arm clip for transcatheter edge-to-edge repair effectively reduced mitral and tricuspid regurgitation and improved hemodynamics without significantly increasing transvalvular pressure gradients.
Long-arm clips feasible ex vivo for complex TEER; leaves open clinical translation pending human studies.
Background A longer-arm clip design for transcatheter edge-to-edge repair of mitral and tricuspid valves could be beneficial in treating complex valve pathologies. Its hemodynamic effects and usability are unknown. This study aims to assess its new design in an ex-vivo beating heart model. Methods The long-arm clip was implanted in porcine left (n = 14) and right (n = 6) hearts with induced degenerative mitral regurgitation and functional tricuspid regurgitation, respectively. Hemodynamic conditions were assessed at baseline, pathology and post-treatment. Usability and grasping quality were evaluated during simulated treatment. Results Mitral valve treatment significantly increased cardiac output ( p < 0.001) and decreased mean left atrial pressure during ventricular systole ( p = 0.001) with respect to pathological conditions. Tricuspid treatment with grasping involving septal leaflet significantly increased cardiac output (posterior-septal grasping: p = 0.006; anterior-septal grasping: p = 0.04). There was no significant increase of transvalvular gradient pressure nor tissue damage. Conclusion Long-arm clip treatment was feasible in porcine hearts, it effectively reduced regurgitation and did not significantly increase the transvalvular pressure gradient. Abbreviations MV, mitral valve; DMR, degenerative mitral regurgitation; FTR, functional tricuspid regurgitation; TV, tricuspid valve; XTR Clip, long-arm clip; P2, posterior middle scallop of mitral valve; MRF, mitral regurgitation fraction; A-P, medial grasping of anterior and posterior leaflets; A-S, medial grasping of anterior and septal leaflets; P-S, medial grasping of posterior and septal leaflets; CO s , systemic cardiac output; CO p , pulmonary cardiac output; AoP, mean aortic pressure; PAP, mean pulmonary artery pressure; LAP, mean left atrial pressure; LAP syst , mean left atrial pressure during ventricular systole; Δp m , mean diastolic pressure gradient across mitral valve; Δp t , mean diastolic pressure gradient across tricuspid valve; APd, mid-systole mitral antero-posterior distance; CI, confidence interval
No takes yet. Share an insight, caveat, or question.
Jaworek et al. (2019) studied this question.