Key result
MitraClip XTR reduced tricuspid annular diameter by ~31% in a severe secondary TR case.
Why the study?
Does the MitraClip XTR device reduce tricuspid annular size and improve symptoms in a patient with severe secondary tricuspid regurgitation?
Case Report (n=1)
Does the MitraClip XTR device reduce tricuspid annular size and improve symptoms in a patient with severe secondary tricuspid regurgitation?
The novel MitraClip XTR device enabled significant tricuspid annular size reduction and symptom improvement in a patient with severe secondary tricuspid regurgitation.
Answers are grounded in this paper and cite their sources.
Supports feasibility of MitraClip XTR for secondary TR annular reduction; hypothesis-generating and requires prospective validation.
The novel MitraClip XTR device has significantly longer clip arms compared with its predecessor to simplify leaflet grasping (Panel A). We applied it to treat a patient with severe secondary tricuspid regurgitation (TR). The patient (female, 57 years) had undergone lung transplantation for idiopathic pulmonary arterial hypertension and subsequent MitraClip NTR implantation for severe mitral regurgitation 8 years ago. Echocardiography revealed mild mitral regurgitation, on invasive assessment mean pulmonary artery pressure was 24 mmHg. She presented with severe symptoms [New York Heart Association (NYHA) Class III] and recurrent peripheral oedema despite maximal diuretic therapy. The XTR clip was placed to connect the septal and the anterior leaflet at the point of maximum regurgitant flow. A second Clip (NTR) was placed next to the first one for stabilization. The tricuspid annular diameter was reduced from 41–43 mm (Panels B and C) to 27–31 mm (Panels F and G). Accordingly, the effective regurgitant orifice area (EROA) decreased from 160 mm2 to 31 mm2 (Panels D and H) and the tricuspid perimeter decreased from 21.1 cm to 12.1 cm (Panels E and I; Supplementary material online, Video 1). At 4-month follow-up, the patient had significantly improved (NYHA Class II and mild TR). The reduction in tricuspid annulus size with the novel XTR device was more than expected and reassuring. While a significant reduction of annular dimensions should provide a better long-term durability, increased tension on the valve leaflets may theoretically impose a higher risk for leaflet tearing: the implantation of a second clip was therefore performed to increase stability. Long-term follow-up data from studies such as the TriValve registry will be required to optimize patient selection. Supplementary data are available at European Heart Journal - Cardiovascular Imaging online.
No takes yet. Share an insight, caveat, or question.
Andreas et al. (2019) conducted a case report in Severe secondary tricuspid regurgitation (n=1). MitraClip XTR was evaluated on Tricuspid annular diameter, effective regurgitant orifice area, and tricuspid perimeter. Transcatheter placement of a MitraClip XTR device in a 57-year-old woman with severe secondary tricuspid regurgitation reduced the tricuspid annular diameter from 41-43 mm to 27-31 mm.
Synapse has enriched 10 related papers. Consider them for comparative context: