Why the study?
Does a Doppler technique accurately measure beat-to-beat cardiac output compared to thermodilution and nitrous oxide rebreathing in humans?
Does a Doppler technique accurately measure beat-to-beat cardiac output compared to thermodilution and nitrous oxide rebreathing in humans?
The Doppler technique combined with M-mode echocardiography provides highly accurate, non-invasive beat-to-beat measurements of cardiac output that correlate excellently with invasive standards.
Answers are grounded in this paper and cite their sources.
May support non-invasive CO monitoring in controlled settings; leaves open broader validation against invasive standards.
We have measured aortic flow in the ascending aorta in man with a Bach-Simpson BVM 202 blood velocity meter, and aortic root diameter by M-mode echocardiography, and thus derived beat-to-beat cardiac output (Q). We tested the technique in 21 patients (53 comparisons) with various cardiovascular problems against a thermodilution method, and in four normal subjects at rest and two levels of exercise (50 and 100 W) against a nitrous oxide rebreathing method. We obtained excellent overall correlation in a range of 0.5-10 litres/min (r = 0.98, n = 77, sy,x = 0.48 litre/min), the formula for the least squares regression being: (Q Doppler) = 0.95 (Q Thermodilution/N2O) + 0.11 litre/min. The Doppler signal is sufficiently noise-free to obtain maximum acceleration of flow from the first derivative of velocity.
No takes yet. Share an insight, caveat, or question.
Mehta et al. (1985) studied this question.