Key result
Prospectively ECG-triggered sequential dual-source CCTA demonstrated high diagnostic accuracy for detecting coronary stenosis in patients with atrial fibrillation, with 89.7% sensitivity and 99.4% specificity on a per-segment basis.
Why the study?
Does prospectively ECG-triggered sequential dual-source CCTA provide diagnostic image quality and accuracy for detecting coronary stenosis in patients with persistent atrial fibrillation regardless of heart rate?
Observational (n=85)
No
Does prospectively ECG-triggered sequential dual-source CCTA provide diagnostic image quality and accuracy for detecting coronary stenosis in patients with persistent atrial fibrillation regardless of heart rate?
Effect estimate: kappa=0.91
Prospectively ECG-triggered sequential dual-source CCTA provides high diagnostic accuracy and image quality in patients with persistent atrial fibrillation, independent of heart rate and heart rate variability.
May support noninvasive coronary evaluation in AF; leaves open randomized validation before practice change.
OBJECTIVES: To evaluate the effects of mean heart rate (HR) and heart rate variation (HRV) on image quality and diagnostic accuracy of prospectively ECG-triggered sequential dual-source coronary CT angiography (CCTA) in patients with atrial fibrillation (AF). METHODS: Eighty-five patients (49 women, 36 men; mean age 62. 1 ± 9.5 years) with persistent AF underwent prospectively ECG-triggered sequential second-generation dual-source CCTA. Tube current and voltage were adjusted according to body mass index (BMI) and iterative reconstruction was used. Image quality of coronary segments (four-point scale) and presence of significant stenosis (>50%) were evaluated. Diagnostic accuracy was analyzed in 30 of the 85 patients who underwent additional invasive coronary angiography (ICA). RESULTS: Only 8 of 1102 (0.7%) segments demonstrated poor image quality. No significant impact on image quality was found for mean HR (94.9 ± 21.8 bpm; r=0.034, p=0.758; F=0.413, p=0.663) or HRV (67.5 ± 22.8 bpm; r=0.097, p=0.377; F=0.111, p=0.895). On per-segment analysis, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were 89.7% (26/29), 99.4% (355/357), 92.9% (26/28), and 99.2% (355/358), respectively, with excellent correlation (kappa=0.91) with ICA. Mean effective dose was 3.3 ± 1.0 mSv. CONCLUSIONS: Prospectively ECG-triggered sequential dual-source CCTA provides diagnostic image quality and good diagnostic accuracy for detection of coronary stenosis in AF patients without significant influence by HR or HRV.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2015) conducted an observational in Persistent Atrial Fibrillation (n=85). Prospectively ECG-triggered sequential dual-source CCTA vs. Invasive coronary angiography (ICA) was evaluated on Diagnostic accuracy for detection of significant coronary stenosis (>50%) on a per-segment basis (kappa=0.91). Prospectively ECG-triggered sequential dual-source CCTA demonstrated high diagnostic accuracy for detecting coronary stenosis in patients with atrial fibrillation, with 89.7% sensitivity and 99.4% specificity on a per-segment basis.