Key result
Guidant's reconfirmation algorithm utilizes non-committed shocks to prevent inappropriate treatment of ventricular tachycardia episodes terminating during or immediately after device charging.
Reviews Guidant's reconfirmation algorithm to aid in troubleshooting inappropriate shocks from implantable cardioverter defibrillators.
Answers are grounded in this paper and cite their sources.
May aid ICD troubleshooting in Guidant devices; leaves open validation in contemporary platforms.
Ventricular tachycardia episodes terminating during or immediately after device charging are not expected to be treated by current implantable cardioverter defibrillators as shocks are non-committed. For proper troubleshooting, the reconfirmation algorithm will be reviewed.
No takes yet. Share an insight, caveat, or question.
Mansour et al. (2009) studied this question. Guidant's reconfirmation algorithm utilizes non-committed shocks to prevent inappropriate treatment of ventricular tachycardia episodes terminating during or immediately after device charging.
Synapse has enriched 10 related papers. Consider them for comparative context: