Why the study?
The effects of sample matrix on high-sensitivity cardiac troponin I concentrations and cutoffs in rapid myocardial infarction algorithms were largely unknown.
Does the use of plasma versus serum sample types cause patient misclassification in 0/1 hour hs-cTnI algorithms?
Does the use of plasma versus serum sample types cause patient misclassification in 0/1 hour hs-cTnI algorithms?
Using different sample types (plasma vs. serum) for hs-cTnI assays can lead to significant patient misclassification in 0/1 hour algorithms, reinforcing the need to use a single sample type.
Answers are grounded in this paper and cite their sources.
EDTA plasma may disrupt troponin complex integrity in assays; extends sample-type observations but leaves standardization needs open.
Over the past 20 years there has been much work in understanding the release and processing of the cardiac troponin complex (troponin C, troponin I, troponin T) in the circulation (1, 2). Of particular interest to clinical laboratories are studies that have demonstrated the impact of sample type and the cardiac troponin complex. Here, ethylenediaminetetraacetic acid (EDTA) has been observed to break up the cardiac troponin complex into the individual subunits with thrombin (i.e., serum samples) cleaving cardiac troponin T (1, 2). During the transition from the contemporary cardiac troponin assays to the high-sensitivity cardiac troponin (hs-cTnI or hs-cTnT) assays, manufacturers have opted to limit the number of sample types listed in their package inserts. For example, for the contemporary cTnI assays, both Siemens ADVIA Centaur and Ortho VITROS list serum, lithium heparin plasma, and EDTA plasma as sample types; however for their respective hs-cTnI assays, only serum and lithium heparin plasma are listed as sample types. Since many clinical studies do not specify their sample types or use more than one sample type in their study and analyses, the effects of matrix on concentrations and cutoffs used in rapid algorithms for myocardial infarction are largely unknown. To this end, we assessed the impact of EDTA or lithium heparin plasma versus serum hs-cTnI concentrations on potential misclassification using the published 0/1 h algorithms (3–5). Briefly, two studies were performed. Study 1 included leftover and de-identified patient samples that had EDTA plasma, lithium heparin plasma, and serum samples collected concurrently, were removed from cells or clot within 8 h of collection, and stored below -20°C. Only matched serum, EDTA, and lithium heparin plasma samples from the same patients that had sufficient volume after thawing (second thaw) for testing for Abbott ARCHITECT hs-cTnI, Siemens ADVIA Centaur hs-cTnI, Beckman Access hs-cTnI, and Ortho VITROS hs-cTnI were included in the study. As the Abbott hs-cTnI assay was the first hs-cTnI assay with the 0/1 h algorithm derived, serum concentrations (n = 11) above and below important cutoffs used in the algorithm (i.e., 2, 5, 52 ng/L) and close to the reported 99th-percentile concentrations (16, 26, 34 ng/L) were selected for comparison against EDTA and lithium heparin plasma for the Abbott, Siemens Centaur, Beckman and Ortho hs-cTnI assays using their respective 0/1 h algorithms (Fig. 1). Study 2 included 165 paired lithium heparin and serum samples obtained (Abbott hs-cTnI serum range = 1.0–48 ng/L, average = 5.4 ng/L), processed and stored as above. For both studies, samples were thawed at room temperature for approximately 45 min, vortexed, and centrifuged (3000 g) for 10 min prior to testing in the following order: Ortho, Beckman, Siemens, Abbott (for study 2 only Ortho and Abbott hs-cTnI measured) with time from aliquoting to placement on analyzers ranging from 0.5 to 2 h. A misclassification (using rounded values to whole number) was noted if either the EDTA plasma or lithium heparin plasma results as compared to the serum results for the individual patient either exceeded the change criteria or misplaced the individual in the rule-out, observe, rule-in arms of the respective 0/1 h algorithms. The differences in concentrations between the plasma (EDTA or lithium heparin) and serum were plotted against the serum hs-cTnI concentrations. Serum was selected as the referent sample type, as this matrix was used in the 3 recent publications on the 0/1 h algorithms for the Siemens Centaur, Beckman, and Ortho hs-cTnI assays (3–5). Difference between plasma (EDTA or lithium heparin) versus serum hs-cTnI concentrations for 4 different hs-cTnI assays (Abbott, Siemens, Beckman, Ortho) (A) and between lithium heparin plasma and serum for 2 different hs-cTnI assays (Abbott, Ortho) (B). For panel A (n = 11 patients) misclassification by using EDTA or lithium heparin plasma versus serum to rule-out using the single low cutoff yielded 1 misclassification for both the Siemens and Beckman hs-cTnI assays. A delta to change classification from rule-out to observe occurred for 1 patient for both the Abbott and Ortho hs-cTnI assays. For rule-in using the single high cutoff, there was 1 misclassification for the Abbott assay and 2 misclassifications for the Ortho hs-cTnI assay with the delta also exceeded (see asterisk). For panel B (n = 165 patients) misclassification by using lithium heparin plasma versus serum to rule-out using the single low cutoff yielded 14 misclassifications for Abbott and 13 misclassifications for Ortho. A delta to change classification from rule-out to observe occurred for 7 patients for Abbott and 9 patients for Ortho. For rule-in using the single high cutoff, there was 1 misclassification for Abbott that did not exceed the delta. For both panels A and B, points marked with an * (asterisk) represent deltas that would misclassify patients to rule-in with either the Abbott or Ortho hs-cTnI assays. Difference between plasma (EDTA or lithium heparin) versus serum hs-cTnI concentrations for 4 different hs-cTnI assays (Abbott, Siemens, Beckman, Ortho) (A) and between lithium heparin plasma and serum for 2 different hs-cTnI assays (Abbott, Ortho) (B). For panel A (n = 11 patients) misclassification by using EDTA or lithium heparin plasma versus serum to rule-out using the single low cutoff yielded 1 misclassification for both the Siemens and Beckman hs-cTnI assays. A delta to change classification from rule-out to observe occurred for 1 patient for both the Abbott and Ortho hs-cTnI assays. For rule-in using the single high cutoff, there was 1 misclassification for the Abbott assay and 2 misclassifications for the Ortho hs-cTnI assay with the delta also exceeded (see asterisk). For panel B (n = 165 patients) misclassification by using lithium heparin plasma versus serum to rule-out using the single low cutoff yielded 14 misclassifications for Abbott and 13 misclassifications for Ortho. A delta to change classification from rule-out to observe occurred for 7 patients for Abbott and 9 patients for Ortho. For rule-in using the single high cutoff, there was 1 misclassification for Abbott that did not exceed the delta. For both panels A and B, points marked with an * (asterisk) represent deltas that would misclassify patients to rule-in with either the Abbott or Ortho hs-cTnI assays. Below 25 ng/L there were minor differences (normally distributed) between EDTA plasma versus serum (average hs-cTnI difference = -0.63 ng/L; 95%CI: -1.30 to 0.04) and between lithium heparin plasma versus serum (average hs-cTnI difference = 0.08 ng/L; 95%CI: -0.24 to 0.41) (Fig. 1). Above 30 ng/L the differences (normally distributed) between EDTA plasma and serum were more evident (average = -6.08 ng/L; 95%CI: -9.53 to -2.62) as compared to lithium heparin plasma and serum (average = -0.13 ng/L; 95%CI: -3.10 to 2.85). Of the 11 patients with 33 measurements per assay, differences between plasma and serum yielded more misclassifications for the 0/1 h algorithms with Abbott hs-cTnI (n = 5) and Ortho hs-cTnI (n = 8) as compared to Beckman hs-cTnI (n = 1) and Siemens hs-cTnI (n = 1) (Fig. 1). Assessing the 0/1 h algorithm in 165 patients with paired lithium heparin and serum samples revealed a similar percentage of misclassifications using Abbott hs-cTnI (15.2%; 95%CI: 9.8 to 22.4) and Ortho hs-cTnI (15.8%; 95%CI: 10.3 to 23.1) assays (Fig. 1). These findings provide evidence that different sample types alone may result in patient misclassification when using the 0/1 h algorithms. This information reinforces laboratory recommendations to accept only one sample type in the course of investigating patients with possible myocardial infarction. Ortho Clinical Diagnostics for funding; we would also like to thank the staff at the Clinical Research Laboratory and Biobank (Hamilton, Ontario) for the management, storage, and testing of the study samples. Ortho does not list EDTA plasma as a recommended sample type in their hsTnI package insert.
No takes yet. Share an insight, caveat, or question.
Kavsak et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: