Key result
The 3-site Ortho hs-cTnI assay demonstrated significantly lower immunoreactivity to macrocomplexes compared with the Abbott hs-cTnI assay in three patients with suspected macrocomplex interference.
Why the study?
Macrocomplexes can cause discrepant hs-cTn concentrations, and whether 3-site immunoassays consistently yield higher immunoreactivity for macro-cTnI remained contradictory.
Does the Ortho 3-site hs-cTnI immunoassay have lower immunoreactivity to macrocomplexes compared to other hs-cTnI assays in patients with suspected falsely elevated troponin?
Case Report (n=3)
Does the Ortho 3-site hs-cTnI immunoassay have lower immunoreactivity to macrocomplexes compared to other hs-cTnI assays in patients with suspected falsely elevated troponin?
The Ortho 3-site hs-cTnI assay may have lower immunoreactivity to macrocomplexes compared to other assays, which can help resolve clinical dilemmas of falsely elevated troponin.
Answers are grounded in this paper and cite their sources.
May aid resolution of suspected macro troponin false positives; hypothesis-generating and requires prospective validation.
The publication by Lam and colleagues (1,) indicates that macrocomplexes (i.e., in vivo antibody–troponin complexes) are important variables that may lead to discrepant concentrations between high-sensitivity cardiac troponin (hs-cTn) assays. Their data on 5 different hs-cTn assays also suggest that the assay design may have a role in this process, with “higher immunoreactivity for macro-cTnI specimens using a 3-site immunoassay (Siemens hs-cTnI Centaur or Siemens hs-cTnI Vista) compared with the other 2-site cTnI immunoassays” (1,). These data are in contrast to the findings by Warner and Marshall (2,), who originally identified a higher rate of macrocomplexes with the Abbott hs-cTnI assay (2-site immunoassay) compared with the contemporary Ortho VITROS Troponin I ES assay, which is a 3-site immunoassay. The IFCC Committee on Clinical Applications of Cardiac Bio-Markers website (https://www.ifcc.org/ifcc-education-division/emd-committees/committee-on-clinical-applications-of-cardiac-bio-markers-c-cb/) provides more details on point-of-care, contemporary, and hs-cTn assays. In 2019, Ortho Clinical Diagnostics gained regulatory approval for its 3-site hs-cTnI immunoassay assay in Canada (VITROS® Immunodiagnostic Products hs Troponin I) (3). The presence of macrocomplexes in patients’ samples poses a dilemma for clinical researchers because these complexes can exist in samples frozen for >15 years; the limited volume of sample obtained in clinical studies often prevents additional testing for interferences (3,). For the treating cardiologist seeking to know what led to an increased hs-cTn result, this is a major dilemma (4,). After obtaining ethics-board approval, we assessed the Ortho hs-cTnI assay in 3 different patients in whom the cardiologist suspected a macrocomplex as a source of the high hs-cTnI concentration shown by the Abbott hs-cTnI assay. The first patient was a 48-year-old woman with a final diagnosis of noncardiac chest discomfort confirmed by coronary computed tomography angiography that showed normal coronary arteries with no evidence of atherosclerotic disease or coronary stenoses. A plasma sample submitted to the laboratory to investigate whether a macrocomplex was present (via polyethylene glycol [PEG] precipitation, as described previously (4,)) yielded a result of 130 ng/L by the Abbott hs-cTnI assay. The hs-cTnI recovery after PEG precipitation was <0.5% (Abbott result, 0.6 ng/L), suggestive of a macrocomplex as the reason for the increased Abbott hs-cTnI result. A plasma aliquot from this patient was further measured on the Siemens Dimension EXL hs-cTnI assay, which was not assessed by Lam and colleagues but is also a 3-site immunoassay (1, 5). The result on the Dimension EXL hs-cTnI assay was 213 ng/L, in agreement with the findings of Lam and colleagues that a 3-site immunoassay may yield higher immunoreactivity. However, subsequent measurement of the aliquot by the Ortho hs-cTnI assay yielded a low result of 4.5 ng/L, indicating that the 3-site immunoassay design by Ortho may not yield higher immunoreactivity for some macrocomplexes. The second patient, a 76-year-old woman whose electrocardiogram demonstrated sinus rhythm with no ischemic changes and who had a normal ejection fraction with no wall motion abnormalities on echocardiography and a normal nuclear perfusion study, exhibited hs-cTnI results over 2 days of 170 ng/L and 174 ng/L by the Abbott hs-cTnI assay. PEG precipitation yielded a recovery of 2.3% (Abbott result, 4.0 ng/L) with an aliquot measured with the Ortho hs-cTnI assay yielding a low result of 3.4 ng/L. The third patient, a 65-year-old woman with a history of remote myocarditis and chronic moderately severe left ventricular dysfunction, presented with cardiogenic shock. During her hospital stay, she had persistently high, nonchanging hs-cTnI results over nearly a week (hs-cTnI range, 6343–6891 ng/L) and similar high values over the preceding several months. Paired EDTA and lithium heparin plasma samples yielded results of 6891 ng/L and 7153 ng/L, respectively, by the Abbott hs-cTnI assay. PEG precipitation of the lithium heparin plasma sample yielded a recovery of 5.3% (Abbott result, 382 ng/L), whereas the Ortho hs-cTnI assay gave a result of 242 ng/L. These results indicate the presence of a macrocomplex, but it is unclear to what extent the high concentration of cTnI was due to injury versus injury and macrocomplex. These case findings indicate that some 3-site immunoassays—specifically, the Ortho hs-cTnI assay—have significantly lower immunoreactivity to macrocomplexes compared with other 3-site and 2-site hs-cTnI immunoassays. Such information may be helpful during the clinical workup of certain patients. It appears that variables besides the number of antibodies used in the immunoassay are important. More thorough investigations with different assays and in different geographic locations will be needed to gain better insight into this evolving area of the effects of endogenous antibodies on hs-cTn assay measurements. Finally, the heterogeneity of these complexes is likely to be very large, and the ability to make generalizations will be difficult. Additional studies using other methodologies for depleting immunoglobulins and biochemically characterizing these complexes will be required.
No takes yet. Share an insight, caveat, or question.
Kavsak et al. (2020) conducted a case report in Suspected macrocomplex interference in hs-cTnI assays (n=3). Ortho hs-cTnI assay vs. Abbott hs-cTnI assay was evaluated on hs-cTnI concentration and recovery after PEG precipitation. The 3-site Ortho hs-cTnI assay demonstrated significantly lower immunoreactivity to macrocomplexes compared with the Abbott hs-cTnI assay in three patients with suspected macrocomplex interference.