Case Report (n=1)
Pacemaker-lead-associated right ventricular perforation is a life-threatening complication that requires urgent intervention in hemodynamically unstable patients.
Answers are grounded in this paper and cite their sources.
May prompt closer post-pacemaker monitoring for early tamponade; single case leaves open incidence and risk-factor questions.
Pacemaker-lead-associated right ventricular perforation is a life-threatening complication. Acute perforation usually presents within 24 hours. Patients with lead perforation are often asymptomatic but fatal complications like hemopericardium, leading to cardiac tamponade and death, are reported. Diagnosis is based on chest x-ray, computed tomography (CT) scan, and echocardiography. The management of the lead perforation is based on clinical presentation. Extraction is avoided in cases of chronic asymptomatic lead perforations because of the associated complications. Urgent intervention is needed in hemodynamically unstable patients with pericardial effusion or cardiac tamponade physiology.
No takes yet. Share an insight, caveat, or question.
Khalid et al. (2018) conducted a case report in Right ventricular perforation post pacemaker insertion complicated with cardiac tamponade (n=1). Dual chamber permanent pacemaker placement was evaluated on Right ventricular perforation and cardiac tamponade.
Synapse has enriched 10 related papers. Consider them for comparative context: