Key result
Electrical interference from a refrigerator triggered an inappropriate implantable cardioverter-defibrillator shock in a 4-year-old child by mimicking ventricular fibrillation.
Household appliances such as refrigerators can cause electrical interference leading to inappropriate ICD shocks in pediatric patients.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to household electromagnetic interference risk in pediatric ICD patients; leaves open broader device vulnerability assessment.
We report a patient with long QT syndrome who received an inappropriate implantable cardioverter-defibrillator shock due to electrical interference from a refrigerator. This electrical interference was mistakenly detected as an episode of ventricular fibrillation and ended with an inappropriate delivery of shock without any warning symptoms before.
No takes yet. Share an insight, caveat, or question.
Sisko et al. (2021) studied this question. Electrical interference from a refrigerator triggered an inappropriate implantable cardioverter-defibrillator shock in a 4-year-old child by mimicking ventricular fibrillation.
Synapse has enriched 10 related papers. Consider them for comparative context: