Key result
A newly designed single circular Teflon pledget ('Ring Pledget') provides secure and uniform shrinkage of the ventricular apex to prevent apical rupture during transapical aortic valve implantation.
Why the study?
Does a single circular Teflon pledget provide secure apex closure and prevent apical rupture during transapical aortic valve implantation?
Case Report (n=42)
Does a single circular Teflon pledget provide secure apex closure and prevent apical rupture during transapical aortic valve implantation?
Answers are grounded in this paper and cite their sources.
A newly designed single circular Teflon pledget may help prevent apical rupture during transapical aortic valve implantation by providing secure and uniform closure of the ventricular apex.
Transapical aortic valve implantation requires puncture of the left ventricle apex and insertion of a 32-French delivery sheath. A critical step in the procedure consists of secure closure of the ventricular apex. We describe 2 cases of apical rupture of 42 transapical aortic valve implantations. Furthermore, we describe the use of a newly designed single circular Teflon pledget that can help to avoid this complication. This pledget provides a more secure and uniform shrinkage of the entire apex to close the defect left by the delivery sheath.
No takes yet. Share an insight, caveat, or question.
Astarci et al. (2010) conducted a case report in Transapical aortic valve implantation (n=42). Single circular Teflon pledget ('Ring Pledget') was evaluated on Apical rupture. A newly designed single circular Teflon pledget ('Ring Pledget') provides secure and uniform shrinkage of the ventricular apex to prevent apical rupture during transapical aortic valve implantation.
Synapse has enriched 10 related papers. Consider them for comparative context: