Key result
SAFEX device successfully seals 21-Fr apical access sites without bleeding or dislodgement in a pig model.
A novel self-expandable apical closure device successfully sealed large-sized apical access sites in a porcine model, even under induced systemic hypertension, supporting its potential for full-percutaneous transapical valve procedures.
Answers are grounded in this paper and cite their sources.
Successful apical sealing in pigs under hypertension; leaves open translation to human transapical procedures.
OBJECTIVES: Transapical valve implantation is traditionally performed through a left antero-lateral mini-thoracotomy. A self-expandable apical closure device has recently been developed for full-percutaneous transapical valve implantation. We performed haemodynamics stress-tests on an animal model to evaluate the sealing properties. METHODS: Under general anaesthesia 5 pigs (mean weight: 67 ± 6 Kg) received full heparinization (100 IU/Kg; activated clotting time >250 s and, through inferior mini-sternotomies, 21-Fr introducer sheaths for transapical aortic valve implantation (outer diameter: 25-Fr) were placed over-the-wire in the apexes. Delivery-catheters carrying folded occluders (SAFEX TM final design) were inserted in the introducer sheaths and plugs were then deployed under fluoroscopic guidance. Phase 1: after protamine injection, apical bleeding was monitored for 1 h with standard haemodynamics condition. Phase 2: we induced systemic hypertension with adrenaline infusion to test the sealing properties under stress. Animals were sacrificed after Phase 2 and hearts were removed and inspected. RESULTS: Five plugs were successfully introduced and deployed in 5 pig hearts. Plugs provided good apical sealing in each animal and a mean of 7 ± 4 ml of blood lost per animal was collected during Phase 1: haemodynamics remained stable and no plug dislodgement was detected (mean blood pressure: 52 ± 9 mmHg). During Phase 2, mean systolic and diastolic peak levels reached 268 ± 24 mmHg and 175 ± 17 mmHg, respectively, without plug dislodgment or bleeding. Post-mortem inspection showed good plug deployment and fixation without myocardial damage. CONCLUSIONS: The new apical occluder seals large-sized apical access sites in animal models also during induced systemic hypertension. This pilot study is a further step towards full-percutaneous transapical valve procedures in the clinical setting.
No takes yet. Share an insight, caveat, or question.
Ferrari et al. (2016) studied Transapical valve implantation (animal model) (n=5). Self-expandable apical closure device (SAFEX TM) was evaluated on Apical sealing properties (blood loss and plug dislodgement). In a pig model, the SAFEX apical closure device successfully sealed 21-Fr apical access sites without dislodgement or bleeding during both standard hemodynamics and induced systemic hypertension.
Synapse has enriched 10 related papers. Consider them for comparative context: