This is the authors' abstract. We don't add key points for this paper.
By mixing hydroxylapatite (HA) into L(-)-dilactide monomer, prior to polymerization to poly(L-lactide) (PLLA), hydroxylapatite filled poly(L-lactide) composites were obtained. This study reports about the mechanical properties of these composites compared with unfilled PLLA. It was concluded that a 30 wt% HA/PLLA composite has better compressive and tensile strengths, higher stiffness and Vickers hardness number than unfilled PLLA (Mv: 125-150,000). Gas sterilization (ethylene oxide) affects molecular weight and flexural strength significantly. Implantation studies revealed loss of 50% of initial flexural strength within 3 weeks, and a faster decline of flexural strength was observed in phosphate buffered saline than in the subcutis of goats. From a mechanical point of view storage at -20 degrees C proved to be a safe method. In its current state HA/PLLA composites can not be used as implant materials that have to resist major forces. However, such composites might be useful in non-loadbearing applications in orthopedic or maxillofacial surgery.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Verheyen et al. (1992) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: