Key result
Transfemoral aortic valve replacement with the Edwards Sapien 3 valve resulted in none/trace aortic regurgitation at 30 days in 89.7% of patients compared to 34.3% with the Edwards Sapien XT valve.
Why the study?
Does the Edwards Sapien 3 valve improve none/trace aortic regurgitation at 30 days compared to the Edwards Sapien XT valve in intermediate-high-risk surgical patients with symptomatic aortic stenosis?
Cohort (n=209)
Does the Edwards Sapien 3 valve improve none/trace aortic regurgitation at 30 days compared to the Edwards Sapien XT valve in intermediate-high-risk surgical patients with symptomatic aortic stenosis?
Answers are grounded in this paper and cite their sources.
The Edwards Sapien 3 valve significantly reduces residual paravalvular leakage at 30 days compared to the older Edwards Sapien XT valve in patients undergoing transfemoral TAVR.
OBJECTIVES: The aim of this study was to compare the 30-day procedural, clinical and echocardiographic outcome of the new balloon-expandable Edwards Sapien 3 (ES3) valve with the Edwards Sapien XT (ESXT). BACKGROUND: Post-implant paravalvular leaks (PVL) after transfemoral aortic valve replacement (TAVR) resulting in residual aortic regurgitation (AR) are a major limitation for long term outcome. New TAVR-devices have to eliminate this problem. METHODS: Transfemoral TAVR was performed in 209 consecutive intermediate-high-risk surgical patients (pts) with symptomatic aortic stenosis (ESXT n = 102, ES3 n = 107). Transthoracic echocardiography (TTE) and 3-dimensional computed tomography were used for valve size selection. Primary endpoint of the study was none/trace AR derived by TTE 30-days after TAVR. RESULTS: All pts underwent successfully TAVR with a combined device success of 100/102 (99%) in ESXT and 107/107 (100%) in ES3 pts. Fluoroscopy time (ESXT 11.8 ± 0.5 min vs. ES3 10.0 ± 0.5 min, P = 0.003) and contrast (ESXT 188.9 ± 5.6 mL vs. ES3 170.4 ± 4.7 mL, P = 0.04) were significantly lower in ES3 patients. 30-day clinical events did not differ. Transvalvular mean pressure gradients were significantly reduced to 7.4 ± 0.8 mmHg after ESXT and to 10.1± 0.4 mmHg after ES3 implantation. After 30 days none/trace AR was found in 34.3% (n = 35) of all ESXT pts in contrast to 89.7% (n = 96) of all ES3 patients. Moderate-to-severe AR was found rarely (ESXT 2.9% vs. ES3 0%, P = 0.073). CONCLUSIONS: Although there was no significant difference in 30 day mortality, the newer ES3 valve reduced significantly residual paravalvular leakage.
No takes yet. Share an insight, caveat, or question.
Bocksch et al. (2016) conducted a cohort in symptomatic aortic stenosis (n=209). Edwards Sapien 3 (ES3) valve vs. Edwards Sapien XT (ESXT) valve was evaluated on none/trace aortic regurgitation derived by TTE 30-days after TAVR. Transfemoral aortic valve replacement with the Edwards Sapien 3 valve resulted in none/trace aortic regurgitation at 30 days in 89.7% of patients compared to 34.3% with the Edwards Sapien XT valve.