Key result
SOTD filtering delivers the highest contrast for distinguishing carotid plaque components in vivo.
Why the study?
The ability to non-invasively assess carotid plaque composition is critical for stroke risk assessment, but the standard SOTD filter for calculating ARFI log(VoA) had not been compared to alternative filtering methods in vivo.
Does the second-order time derivative (SOTD) filter improve the delineation of carotid plaque components in in vivo ARFI log(VoA) ultrasound imaging compared to PCA, FIR, IIR, and MCS filtering methods in patients undergoing carotid endarterectomy?
Observational (n=17)
No
Does the second-order time derivative (SOTD) filter improve the delineation of carotid plaque components in in vivo ARFI log(VoA) ultrasound imaging compared to PCA, FIR, IIR, and MCS filtering methods in patients undergoing carotid endarterectomy?
The second-order time derivative (SOTD) filter remains the optimal signal processing method for noninvasive in vivo ARFI log(VoA) ultrasound delineation of vulnerable carotid atherosclerotic plaque features.
Answers are grounded in this paper and cite their sources.
Supports SOTD for plaque imaging; leaves open prospective validation before clinical adoption.
Carotid atherosclerosis is a major cause of ischemic stroke, and the ability to non-invasively assess plaque composition and structure is critical to effective stroke risk assessment. Carotid plaque components are delineated noninvasively by Acoustic Radiation Force Impulse (ARFI)-derived Variance of Acceleration, evaluated as its decadic log (log(VoA)). To date, this log(VoA) parameter has been calculated by isolating the variance in ARFI-induced displacement profiles using the second-order time derivative (SOTD), a high-pass filtering operation. The purpose of this study was to compare the performance of the SOTD filter to various other filtering methods in application to delineating human carotid plaque components, in vivo. Specifically, the SOTD filter was compared to Principal Component Analysis (PCA), Finite Impulse Response (FIR), Infinite Impulse Response (IIR), and mean-center spatial (MCS) filters. Filter performances were evaluated in terms of the resulting log(VoA) generalized contrast-to-noise ratio (gCNR) for distinguishing plaque features in human carotid plaques, in vivo, which were validated by spatially aligned histology. Results indicated that the SOTD filter consistently provided the highest gCNR for most plaque components, whereas the performances yielded by the other filters were more variable. The study demonstrated that the SOTD filter remains the preferred method for log(VoA) calculation due to its effectiveness for delineating carotid plaque features.
No takes yet. Share an insight, caveat, or question.
Qazi et al. (2025) conducted an observational in Carotid atherosclerosis (n=17). Second-order time derivative (SOTD) filter for ARFI log(VoA) ultrasound imaging vs. Principal Component Analysis (PCA), Finite Impulse Response (FIR), Infinite Impulse Response (IIR), and mean-center spatial (MCS) filters was evaluated on Generalized contrast-to-noise ratio (gCNR) for distinguishing carotid plaque components (calcium, collagen, lipid-rich necrotic core, and intraplaque hemorrhage) validated by spatially aligned histology (p=<0.05). The second-order time derivative (SOTD) filter consistently yielded the highest generalized contrast-to-noise ratio (gCNR) for discriminating carotid plaque components in vivo, particularly calcium from collagen.
Synapse has enriched 10 related papers. Consider them for comparative context: