Key result
Circadian and circaseptan variations are prevalent components of thoracic impedance, inducing changes of several Ohms that can be an error source for daily measurements.
Why the study?
Are there circadian and circaseptan rhythms in thoracic impedance measured by implantable cardioverter-defibrillators?
Observational (n=53)
Are there circadian and circaseptan rhythms in thoracic impedance measured by implantable cardioverter-defibrillators?
Circadian and circaseptan variations are prevalent in thoracic impedance measurements and can induce changes of several Ohms, potentially acting as an error source for daily fluid load monitoring.
Answers are grounded in this paper and cite their sources.
Requires caution interpreting daily ICD impedance trends; hypothesis-generating for time-adjusted fluid monitoring algorithms.
Thoracic impedance (TI) denotes the electrical resistance that is measured between the right ventricular lead and the can of an implantable cardioverter-defibrillator. It is used in medical diagnostics for fluid load monitoring in heart failure patients. We analysed TI for diurnal and weekly components that go along with the recurrent patterns of daily life.TI was measured in 53 patients over 280 d each with a resolution of 1 h. Information about the patients' professional occupation and typical sleeping rhythm was sought. Periodic signal components were identified in the amplitude spectrum, and their main characteristics were derived by cosinor analysis. The typical daily and weekly pattern were reconstructed by Fourier filtering.The Fourier spectrum indicates a strong circadian and a weaker circaseptan component in TI. The latter is significantly elevated in patients who regularly go to work, on error level [Formula: see text]. Cosinor analysis states a significant circadian rhythm in all patients ([Formula: see text]), with MESOR 61.8 [Formula: see text], amplitude 1.9 [Formula: see text], and acrophase 17.5 h. A significant weekly rhythm is found in 25 patients ([Formula: see text]), with amplitude 0.31 [Formula: see text] and acrophase typically on Wednesday. Both rhythms typically obtain their maximum during the 'active' phase of the period in daily life, i.e. in the afternoon and on working days, respectively, while the minimum is reached in the 'recreative' phase.Circadian and circaseptan variation are hence prevalent components of TI, which can induce impedance changes of several Ohms and thus be an error source for daily TI measurements.
No takes yet. Share an insight, caveat, or question.
Kirchner et al. (2015) conducted an observational in Heart failure / Implantable cardioverter-defibrillator recipients (n=53). Circadian and circaseptan rhythms was evaluated on Circadian and circaseptan components in thoracic impedance. Circadian and circaseptan variations are prevalent components of thoracic impedance, inducing changes of several Ohms that can be an error source for daily measurements.
Synapse has enriched 10 related papers. Consider them for comparative context: