Key result
A novel measure of intrathoracic impedance, DeltaTI, was negative preceding 66% of ventricular tachyarrhythmia episodes and below zero.
Why the study?
Are ventricular tachyarrhythmia (VT/VF) episodes preceded by decreases in intrathoracic impedance in patients with implanted defibrillators?
Observational (n=121)
Are ventricular tachyarrhythmia (VT/VF) episodes preceded by decreases in intrathoracic impedance in patients with implanted defibrillators?
Decreases in intrathoracic impedance, which may reflect fluid accumulation, frequently precede ventricular tachyarrhythmia episodes in patients with implanted defibrillators.
Answers are grounded in this paper and cite their sources.
DeltaTI drops precede most VT/VF; hypothesis-generating for impedance monitoring in ICD patients and requires prospective validation.
BACKGROUND: Heart failure is associated with ventricular tachyarrhythmias (VT/VF). Fluid accumulation during worsened heart failure may trigger VT/VF. Increased intrathoracic impedance has been correlated with fluid accumulation during heart failure. Implanted defibrillators capable of daily measures of intrathoracic impedance allow correlation of impedance with occurrence of VT/VF. We hypothesized that VT/VF episodes are preceded by decreases in intrathoracic impedance. The goal was to identify the relationship of intrathoracic impedance measured by implanted cardioverter defibrillators to the occurrence of VT/VF. METHOD: Implanted defibrillator follow-up data were obtained retrospectively. Those with Medtronic OptiVol (Medtronic Inc., Minneapolis, MN, USA), storing averaged daily and reference impedance values, were reviewed for VT/VF episodes. Impedance changes in the week leading up to VT/VF were analyzed. RESULTS: A total of 317 VT/VF episodes in a cohort of 121 patients' follow-up data were evaluated. Averaged daily intrathoracic impedance declined preceding 64% of VT/VF episodes, with an average decline of 0.46 +/- 0.35 Ohms from the day before the VT/VF episodes. However, the mean values of the averaged daily and reference impedance did not change significantly. A novel measure, DeltaTI, the sum of the daily differences between the averaged daily and reference impedance, was negative preceding 66% of VT/VF episodes (P < 0.001). The mean DeltaTI was -4.0 +/- 1.3 Ohms, which was significantly lower than the theoretically expected value of zero Ohms (P < 0.01). CONCLUSION: (1) Averaged daily impedance declined preceding 64% of VT/VF episodes, but the overall decline was of small magnitude; (2) a novel measure, DeltaTI, was negative preceding 66% of VT/VF episodes, and significantly below zero.
No takes yet. Share an insight, caveat, or question.
Moore et al. (2010) conducted an observational in Heart failure with ventricular tachyarrhythmias (n=121). Intrathoracic impedance changes was evaluated on Intrathoracic impedance changes (DeltaTI) preceding VT/VF episodes (p=<0.001). A novel measure of intrathoracic impedance, DeltaTI, was negative preceding 66% of ventricular tachyarrhythmia episodes and below zero.
Synapse has enriched 10 related papers. Consider them for comparative context: