Answers are grounded in this paper and cite their sources.
Hemofiltration (HF) is a group of blood purification therapies used to treat patients with kidney injury. HF works using a process called ultrafiltration (UF) that removes excess liquid accumulated in the patient’s body caused by lack of excretion. UF progress is monitored by the HF machine, but the state-of-the-art method is cumbersome and could be more accurate. In this paper, a system composed by two optical sensors is proposed for real-time non-invasive estimation of ultrafiltration rate. This new system is simple, rugged, and low-cost and operates on sound theoretical foundations. The sensor system has been tested with two different experimental protocols and showed good correlation between its output and the reference value of the ultrafiltration rate (R2= 0.97), as well as improved accuracy compared with the available commercial machine (12ml/h). This system also has the potential to simplify the architecture required by critical care blood purification machines to perform UF control.
No takes yet. Share an insight, caveat, or question.
Visotti et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: