Why the study?
Does a smartphone heart monitor accurately measure QTc intervals compared to a standard 12-lead ECG in healthy volunteers and patients receiving QT-prolonging antiarrhythmics?
Does a smartphone heart monitor accurately measure QTc intervals compared to a standard 12-lead ECG in healthy volunteers and patients receiving QT-prolonging antiarrhythmics?
A smartphone heart monitor provides accurate QTc measurements compared to a 12-lead ECG in healthy volunteers and can detect QTc prolongation in patients on antiarrhythmics, though accuracy decreases when QTc > 500 ms.
Answers are grounded in this paper and cite their sources.
SHM QTc may support monitoring in healthy volunteers but warrants caution when QTc >500 ms; leaves open prospective validation in antiarrhythmic patients.
INTRODUCTION: A variety of medications ranging from antiarrhythmics to psychotropics, as well as conditions such as bradycardia, can prolong the QT interval, increasing the risk for life-threatening arrhythmias. Monitoring the corrected QT interval (QTc) is therefore critical for patient safety. The recent development of smart phone heart monitors (SHM) may allow for easier QTc monitoring. We sought to evaluate the accuracy of an SHM for assessing the QTc, as compared to the standard 12-lead ECG. METHODS AND RESULTS: We compared the QTc interval in lead-I and lead-II between an SHM and 12-lead ECG. Healthy volunteers and hospitalized patients in sinus rhythm being loaded on dofetilide or sotalol were included. Manual and automatic measurements were studied. Across 99 healthy volunteers, the SHM QTc demonstrated good agreement (bias = 4 milliseconds, standard deviation of bias = 11 milliseconds) compared to the 12-lead ECG, using the Bland-Altman method of agreement. Across all hospitalized patients, the SHM was capable of demonstrating QTc prolongation. Between the 12-lead ECG and SHM, lead-I measurements had reasonable agreement (bias = 3 milliseconds, standard deviation of bias = 46 milliseconds). A QTc of > 500 milliseconds was associated with a higher likelihood (OR = 12.0; 95% CI 1.5-111.4; P = 0.02) to not achieve perfect agreement. CONCLUSION: The SHM is accurate in measuring QTc interval in sinus rhythm when compared to 12-lead ECG in healthy volunteers. For patients receiving QT prolonging antiarrhythmics, SHM is capable of detecting QTc prolongation, and lead-I of the SHM is most accurate in measuring the QTc if < 500 milliseconds.
No takes yet. Share an insight, caveat, or question.
Garabelli et al. (2016) studied this question.