Key result
Pocket-size echographs (Vscan) demonstrated excellent agreement with high-end echocardiographs for left ventricular systolic function and pericardial effusion.
Why the study?
Does a pocket-size handheld echocardiograph provide comparable diagnostic accuracy to a high-end standard echocardiograph in patients requiring urgent echocardiography?
Cross-Sectional (n=104)
Blinded
Does a pocket-size handheld echocardiograph provide comparable diagnostic accuracy to a high-end standard echocardiograph in patients requiring urgent echocardiography?
Pocket-size handheld echocardiographs used by trained cardiologists offer good diagnostic accuracy for key parameters in the acute care setting compared to standard high-end machines.
Answers are grounded in this paper and cite their sources.
May facilitate urgent bedside echocardiography; leaves open whether accuracy improves outcomes in prospective studies.
AIMS: Pocket-size echographs may be useful for bedside diagnosis in acute cardiac care, but their diagnostic accuracy in this setting has not been well tested. Our aim was to evaluate this tool in patients requiring an urgent echocardiogram. METHODS: Trained cardiologists performed echocardiograms with a pocket-size echograph (Vscan) in consecutive patients requiring urgent echocardiography. The exams were then compared in a blinded manner with echocardiograms performed with a high-end standard echocardiograph. RESULTS: A total of 104 patients were studied. There was an excellent agreement between the Vscan and the high-end echocardiograph for the left ventricular systolic function and pericardial effusion (Kappa: 0.89 and 0.81, respectively), and the agreement was good or moderate for evaluating the aortic, mitral, and tricuspid valve function and the left ventricular size (Kappa: 0.55-0.66). Visualization of the Vscan images in full-screen format on a PC did not in general confer added value. CONCLUSION: The Vscan used by a trained cardiologist has good diagnostic accuracy in the emergency setting compared with a high-end echocardiograph, despite small screen size and lack of pulse-wave and continuous Doppler.
No takes yet. Share an insight, caveat, or question.
Testuz et al. (2012) conducted a cross-sectional in Acute cardiac care requiring urgent echocardiogram (n=104). Pocket-size echograph (Vscan) vs. High-end standard echocardiograph was evaluated on Agreement for left ventricular systolic function and pericardial effusion (Kappa: 0.89 and 0.81). Pocket-size echographs (Vscan) demonstrated excellent agreement with high-end echocardiographs for left ventricular systolic function and pericardial effusion.
Synapse has enriched 10 related papers. Consider them for comparative context: