Key result
The FF680 electronic blood pressure monitor met ISO 81060-2:2018 accuracy standards, demonstrating mean differences of -0.66 mmHg for SBP and -0.28 mmHg for DBP.
Why the study?
To evaluate the accuracy of blood pressure measurement in adults using FF680 electronic blood pressure monitors with the Korotkoff-Sound method according to ISO 81060-2:2018 standards.
Does the FF680 electronic blood pressure monitor accurately measure blood pressure in adults according to ISO 81060-2:2018 standards?
Cross-Sectional (n=85)
No
Does the FF680 electronic blood pressure monitor accurately measure blood pressure in adults according to ISO 81060-2:2018 standards?
Mean Difference: -0.66
The FF680 electronic blood pressure monitor meets ISO 81060-2:2018 accuracy standards and is recommended for adult blood pressure monitoring.
FF680 meets ISO accuracy criteria; leaves open need for outcome studies before routine clinical adoption.
This study evaluated the accuracy of blood pressure measurement in adults using FF680 electronic blood pressure monitors with the Korotkoff-Sound method, in accordance with the ISO 81060-2:2018. The study was conducted at Shijiazhuang People's Hospital and lasted 56 days, from 26 May to 21 July 2023. Participants, drawn from an adult demographic, underwent data verification and analysis with strict adherence to the trial protocol. For the FF680 electronic blood pressure monitor using the Korotkoff-Sound method, data from 85 valid participants were analyzed. The findings revealed mean differences (SDs) of -0.66 mmHg (2.45 mmHg) for SBP and -0.28 mmHg (2.09 mmHg) for DBP. SBP and DBP had SDs of ≤6.9 and ≤6.95 mmHg, meeting the standard requirements. FF680 devices are recommended for adult blood pressure monitoring because they meet the ISO 81060-2:2018.
No takes yet. Share an insight, caveat, or question.
Hao Chen (2024) reported a cross-sectional. FF680 electronic blood pressure monitor was evaluated on Mean difference in systolic and diastolic blood pressure (MD -0.66 mmHg for SBP and -0.28 mmHg for DBP). The FF680 electronic blood pressure monitor met ISO 81060-2:2018 accuracy standards, demonstrating mean differences of -0.66 mmHg for SBP and -0.28 mmHg for DBP.