Key result
The Hanvon FY730 upper-arm blood pressure monitor demonstrated high accuracy compared to reference measurements, with mean differences of -0.37 mmHg for SBP and -0.17 mmHg for DBP.
Why the study?
The study was conducted to evaluate the measurement accuracy of the Hanvon FY730 upper-arm blood pressure monitor according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018).
Does the Hanvon FY730 upper-arm blood pressure monitor provide accurate measurements compared to reference methods in adults?
Does the Hanvon FY730 upper-arm blood pressure monitor provide accurate measurements compared to reference methods in adults?
Mean Difference: -0.37
The Hanvon FY730 upper-arm blood pressure monitor meets the AAMI/ESH/ISO universal standard for accuracy, demonstrating its suitability for clinical and home BP monitoring.
May support clinical use of Korotkoff-based monitors; leaves open independent validation in broader populations.
This study evaluates the measurement accuracy of the Hanvon FY730 upper-arm blood pressure (BP) monitor, which uses the Korotkoff sound method, in compliance with the AAMI/ESH/ISO universal standard (ISO 81060-2:2018). A clinical trial involving 86 eligible adults was conducted, and data were analyzed based on standardized protocols. The mean differences between the device and reference measurements for systolic blood pressure (SBP) and diastolic blood pressure (DBP) were -0.37 mmHg (SD: 2.25 mmHg) and -0.17 mmHg (SD: 2.02 mmHg), respectively. Standard deviations for SBP (≤6.93 mmHg) and DBP (≤6.95 mmHg) met the required thresholds. A total of 258 effective paired measurements demonstrated that the Hanvon FY730 provides highly consistent results compared with reference methods, with all deviations falling within acceptable limits. The study also confirmed the device's reliability across a wide range of arm circumferences (22-36 cm) and BP levels (SBP: 90-188 mmHg, DBP: 53-107 mmHg). These findings confirm that the Hanvon FY730 complies with ISO 81060-2:2018 standards, demonstrating its suitability for accurate BP monitoring in clinical, personal, and home settings.
No takes yet. Share an insight, caveat, or question.
Hao Chen (2025) studied this question. Hanvon FY730 upper-arm blood pressure monitor vs. Reference measurements was evaluated on Mean difference between device and reference measurements for systolic and diastolic blood pressure (MD -0.37 mmHg for SBP and -0.17 mmHg for DBP). The Hanvon FY730 upper-arm blood pressure monitor demonstrated high accuracy compared to reference measurements, with mean differences of -0.37 mmHg for SBP and -0.17 mmHg for DBP.