Key result
YuWell YE660D monitor meets AAMI/ESH/ISO validation standards within ~2 mmHg of reference.
Why the study?
The accuracy of the YuWell YE660D oscillometric upper-arm BP monitor for clinic and home use in adults needed evaluation according to the AAMI/ESH/ISO Universal Standard and its Amendment 1. 2020.
Cross-Sectional (n=85)
Answers are grounded in this paper and cite their sources.
Supports YuWell YE660D for clinic/home BP use in adults; leaves open need for independent real-world validation.
OBJECTIVE: To evaluate the accuracy of the YuWell YE660D oscillometric upper-arm blood pressure (BP) monitor in general population (for clinic and home BP measurements in adults) according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2 : 2018) and its Amendment 1. 2020. METHODS: Subjects were recruited to fulfill the age, sex, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for arm circumferences 22-32 cm (standard) and 22-45 cm (wide range). RESULTS: Ninety-two subjects were recruited and 85 subjects were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 0.3 ± 7.2/2.2 ± 5.5 mmHg (systolic/diastolic). For validation criterion 2, the SD of the averaged BP differences between the test device and reference BP per subject was 6.1/4.8 mmHg (systolic/diastolic). CONCLUSION: The YuWell YE660D oscillometric upper-arm electronic BP monitor has passed the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018) and its Amendment 1. 2020 in adults and hence can be recommended for home and clinical use.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2023) conducted a cross-sectional in General population (n=85). YuWell YE660D oscillometric upper-arm blood pressure monitor vs. Reference BP readings was evaluated on Mean difference between test device and reference BP readings (validation criterion 1) (MD 0.3 ± 7.2 mmHg (systolic) / 2.2 ± 5.5 mmHg (diastolic)). The YuWell YE660D blood pressure monitor passed AAMI/ESH/ISO validation criteria with mean differences versus reference of 0.3 ± 7.2 mmHg for systolic and 2.2 ± 5.5 mmHg for diastolic BP.
Synapse has enriched 10 related papers. Consider them for comparative context: