Key result
YuWell YE990 monitor meets international accuracy standards vs mercury sphygmomanometer with mean differences under ~1 mmHg.
Why the study?
The accuracy of the YuWell YE990 automated oscillometric upper-arm blood pressure monitor needed validation according to the AAMI/ESH/ISO Universal standards.
Does the YuWell YE990 blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
Cross-Sectional (n=85)
Does the YuWell YE990 blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in adults?
The YuWell YE990 automated blood pressure monitor meets universal accuracy standards and is recommended for clinical use.
Answers are grounded in this paper and cite their sources.
May support YE990 adoption for routine BP measurement; leaves open need for broader independent validation across populations.
To validate the accuracy of the YuWell YE990 automated oscillometric upper-arm medical blood pressure (BP) monitor in adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal standards (ISO 81060-2:2018 and Amendment 1:2020). Participants were recruited to meet the age, sex, BP, and cuff distribution criteria of the AAMI/ESH/ISO standards. BP was measured using a mercury sphygmomanometer (reference device) and YE990 (test device) following a same-arm sequential protocol with two trained observers and a supervisor. Bland-Altman plots were used to assess agreement, and scatter plots were used to assess the performance across varying arm sizes. Ninety-nine subjects were recruited and 85 were analyzed. The YE990 passed the accuracy requirements of the AAMI/ESH/ISO standards, with mean differences of 0.4 ± 6.31 mmHg for systolic BP and -0.9 ± 6.28 mmHg for diastolic BP for validation criterion 1. For validation criterion 2, the SD of the average BP difference between the test device and reference BP per subject was 5.00/5.70 mmHg (systolic/diastolic). YuWell YE990 meets the AAMI/ESH/ISO universal accuracy standard (ISO 81060-2:2018+Amd.1:2020) and is recommended for clinical use.
No takes yet. Share an insight, caveat, or question.
Xu et al. (2025) conducted a cross-sectional in Blood pressure measurement (n=85). YuWell YE990 automated oscillometric upper-arm medical blood pressure monitor vs. Mercury sphygmomanometer was evaluated on Mean difference in systolic and diastolic blood pressure (validation criterion 1) (MD 0.4 ± 6.31 mmHg (systolic) and -0.9 ± 6.28 mmHg (diastolic)). The YuWell YE990 blood pressure monitor met AAMI/ESH/ISO accuracy standards compared to a mercury sphygmomanometer, with mean differences of 0.4 ± 6.31 mmHg (systolic) and -0.9 ± 6.28 mmHg (diastolic).