Key result
The YuWell YE690A blood pressure monitor met ESH validation criteria, with 79 of 99 systolic and 72 of 99 diastolic measurements within 5 mmHg of mercury sphygmomanometer readings.
Why the study?
Does the YuWell YE690A upper-arm blood pressure monitor provide accurate blood pressure measurements compared to a mercury sphygmomanometer in adults?
Observational (n=33)
Does the YuWell YE690A upper-arm blood pressure monitor provide accurate blood pressure measurements compared to a mercury sphygmomanometer in adults?
The YuWell YE690A upper-arm blood pressure monitor is accurate and meets the European Society of Hypertension International Protocol revision 2010 requirements for adult use.
YE690A meets ESH-IP 2010 criteria in Chinese adults; leaves open need for independent validation across populations before routine use.
OBJECTIVE: The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure monitor YuWell YE690A for blood pressure measurement according to the International Protocol of the European Society of Hypertension revision 2010. PARTICIPANTS AND METHODS: Systolic and diastolic blood pressures were measured sequentially in 33 adult Chinese (12 women, 44.2 years of mean age) using a mercury sphygmomanometer (two observers) and the YE690A device (one supervisor). A total of 99 pairs of comparisons were obtained from 33 participants for judgments in two parts with three grading phases. RESULTS: All the blood pressure requirements were fulfilled. The YuWell YE690A device achieved the targets in part 1 of the validation study. The number of absolute differences between device and observers within 5, 10, and 15 mmHg was 79/99, 96/99, and 97/99, respectively, for systolic blood pressure and 72/99, 95/99, and 98/99, respectively, for diastolic blood pressure. The device also fulfilled the criteria in part 2 of the validation study. Thirty-one and 25 participants for systolic and diastolic blood pressure, respectively, had at least two of the three device-observer differences within 5 mmHg (required ≥24). No participant for systolic and two participants for diastolic blood pressure had all the three device-observer comparisons greater than 5 mmHg. CONCLUSION: The YuWell blood pressure monitor YE690A has passed the requirements of the International Protocol revision 2010 and hence can be recommended for blood pressure measurement in adults.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2017) conducted an observational in Blood pressure measurement (n=33). YuWell YE690A automated oscillometric upper-arm blood pressure monitor vs. Mercury sphygmomanometer was evaluated on Device-observer blood pressure differences within 5, 10, and 15 mmHg. The YuWell YE690A blood pressure monitor met ESH validation criteria, with 79 of 99 systolic and 72 of 99 diastolic measurements within 5 mmHg of mercury sphygmomanometer readings.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: