Why the study?
Do the Microlife BP 3AS1-2 and Omron HEM-SOLAR devices accurately measure blood pressure compared to a conventional Mercury sphygmomanometer in individuals in low-resource settings?
Do the Microlife BP 3AS1-2 and Omron HEM-SOLAR devices accurately measure blood pressure compared to a conventional Mercury sphygmomanometer in individuals in low-resource settings?
Both the Omron HEM-SOLAR and Microlife BP 3AS1-2 devices performed adequately under field conditions in a low-resource setting, though the Omron device was more accurate while the Microlife device was more accepted by healthcare workers.
Answers are grounded in this paper and cite their sources.
These devices may suit low-resource BP screening; leaves open need for prospective outcome trials before routine adoption.
INTRODUCTION: High blood pressure (BP) is known to be the greatest modifiable risk factor of cardiovascular diseases, of which 80% occur in low-resource and middle-resource settings. Yet, BP measurement in these countries remains extremely poor. In 2005, a WHO committee invited manufacturers to produce devices especially for use in low-resource settings. Our study assesses the accuracy, performance, and acceptability of two submitted oscillometric devices under field circumstances in Suriname (South America), namely, the Microlife BP 3AS1-2 and Omron HEM-SOLAR. METHODS: We compared BP measurements of test devices using a conventional Mercury sphygmomanometer, performed by local healthcare workers under field circumstances. Three hundred and forty-two individuals were included. Statistics included t-tests, analysis of variance, and Bland-Altman plots. RESULTS: The mean systolic/diastolic BP differences (SD) were -3.5 (8.0)/-7.0 (6.0) for Omron versus Mercury and -6.4 (7.8)/-6.5 (6.0) mmHg for Microlife versus Mercury. Microlife was more accepted by healthcare workers, and both devices performed adequately under field conditions. DISCUSSION AND CONCLUSION: The acceptability, durability, and performance of both test devices were adequate. However, Microlife BP 3AS1-2 underestimated systolic pressure almost twice as much compared with Omron HEM-SOLAR, with identical diastolic underestimations. Guaranteed global availability, users could make a choice between the Omron HEM-SOLAR being more accurate in BP measurement, and Microlife BP 3AS1-2 being more accepted by healthcare workers.
No takes yet. Share an insight, caveat, or question.
Kewalbansing et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: